Rest and exercise hemodynamics of 20 to 23 mm allograft, Medtronic Intact (porcine), and St. Jude Medical valves in the aortic position.

作者信息

Jaffe W M, Coverdale H A, Roche A H, Whitlock R M, Neutze J M, Barratt-Boyes B G

机构信息

Department of Cardiology, Green Lane Hospital, Auckland, New Zealand.

出版信息

J Thorac Cardiovasc Surg. 1990 Aug;100(2):167-74.

PMID:2385114
Abstract

Doppler echocardiography was used to measure gradients and valve areas at rest and after supine bicycle exercise in 35 patients with valve replacements 20 to 23 mm in size. Thirteen patients with a St. Jude Medical valve (St. Jude Medical, Inc., St. Paul, Minn.) were matched to 13 patients with an allograft valve, and seven patients with a Medtronic Intact (porcine) valve (Medtronic, Inc., Minneapolis, Minn.) to seven patients with an allograft valve. Patients were matched for age, sex, valve size, body surface area, and left ventricular systolic function. There was no statistically significant difference between the matched groups for body surface areas, resting cardiac output, exercise heart rate, or workload achieved. Mean pressure gradient was higher for St. Jude Medical than for allograft groups, both at rest (11.8 +/- 6.67 mm Hg for St. Jude Medical versus 6.67 +/- 2.98 mm Hg for allografts) and after exercise (16.4 +/- 8.47 mm Hg versus 9.7 +/- 3.94 mm Hg), but the differences were of borderline significance (p = 0.016 and 0.027, respectively). Valve area at rest was similar for both devices (1.4 +/- 0.45 cm2 for St. Jude Medical versus 1.8 +/- 0.56 cm2 for allograft; p greater than 0.1). There were highly significant differences between patients with Intact and those with allograft valves for resting mean pressure gradient (19.3 +/- 4.23 mm Hg for Intact versus 5.9 +/- 3.68 mm Hg for allograft; p less than 0.001) and for exercise mean pressure gradient (27.8 +/- 8.63 mm Hg versus 8.1 +/- 8.43 mm Hg; p less than 0.001). The differences between the valve areas at rest also were significant (1.1 +/- 0.12 cm2 versus 2.2 +/- 0.62 cm2; p less than 0.01). It is concluded that when a tissue valve is indicated in patients with a small aortic root, the freehand allograft aortic valve is an ideal device from the hemodynamic perspective and is superior to the Intact valve. It is also probably superior in this respect to the St. Jude Medical valve, although the analysis may be biased slightly in favor of the allograft valve.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验