Rest and exercise hemodynamics of 20 to 23 mm allograft, Medtronic Intact (porcine), and St. Jude Medical valves in the aortic position.

作者信息

Jaffe W M, Coverdale H A, Roche A H, Whitlock R M, Neutze J M, Barratt-Boyes B G

机构信息

Department of Cardiology, Green Lane Hospital, Auckland, New Zealand.

出版信息

J Thorac Cardiovasc Surg. 1990 Aug;100(2):167-74.

PMID:2385114
Abstract

Doppler echocardiography was used to measure gradients and valve areas at rest and after supine bicycle exercise in 35 patients with valve replacements 20 to 23 mm in size. Thirteen patients with a St. Jude Medical valve (St. Jude Medical, Inc., St. Paul, Minn.) were matched to 13 patients with an allograft valve, and seven patients with a Medtronic Intact (porcine) valve (Medtronic, Inc., Minneapolis, Minn.) to seven patients with an allograft valve. Patients were matched for age, sex, valve size, body surface area, and left ventricular systolic function. There was no statistically significant difference between the matched groups for body surface areas, resting cardiac output, exercise heart rate, or workload achieved. Mean pressure gradient was higher for St. Jude Medical than for allograft groups, both at rest (11.8 +/- 6.67 mm Hg for St. Jude Medical versus 6.67 +/- 2.98 mm Hg for allografts) and after exercise (16.4 +/- 8.47 mm Hg versus 9.7 +/- 3.94 mm Hg), but the differences were of borderline significance (p = 0.016 and 0.027, respectively). Valve area at rest was similar for both devices (1.4 +/- 0.45 cm2 for St. Jude Medical versus 1.8 +/- 0.56 cm2 for allograft; p greater than 0.1). There were highly significant differences between patients with Intact and those with allograft valves for resting mean pressure gradient (19.3 +/- 4.23 mm Hg for Intact versus 5.9 +/- 3.68 mm Hg for allograft; p less than 0.001) and for exercise mean pressure gradient (27.8 +/- 8.63 mm Hg versus 8.1 +/- 8.43 mm Hg; p less than 0.001). The differences between the valve areas at rest also were significant (1.1 +/- 0.12 cm2 versus 2.2 +/- 0.62 cm2; p less than 0.01). It is concluded that when a tissue valve is indicated in patients with a small aortic root, the freehand allograft aortic valve is an ideal device from the hemodynamic perspective and is superior to the Intact valve. It is also probably superior in this respect to the St. Jude Medical valve, although the analysis may be biased slightly in favor of the allograft valve.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验