Ball Christopher P, Levick Andrew P, Woolliams Emma R, Green Paul D, Dury Martin R, Winkler Rainer, Deadman Andrew J, Fox Nigel P, King Martin D
Department of Earth Sciences, Royal Holloway University of London, Egham, Surrey, UK.
Appl Opt. 2013 Jul 10;52(20):4806-12. doi: 10.1364/AO.52.004806.
Sintered polytetrafluoroethylene (PTFE) is highly reflective and is widely used as a reference standard in remote sensing, radiometry, and spectroscopy. The relative change in output flux from a PTFE integrating sphere over the room temperature phase transition at 19°C has been measured at a monochromatic wavelength of 633 nm as 1.82±0.21%. The change in output flux was attributed to a small change of 0.09±0.02% in the total hemispherical reflectance of PTFE, caused by a change in its material density as a result of the phase transition. For the majority of users, this small change measured in total hemispherical reflectance is unlikely to impact significantly the accuracy of PTFE flat panel reflectors used as reference standards. However, owing to the multiple reflections that occur inside an integrating sphere cavity, the effect is multiplied and remedial action should be applied, either via a mathematical correction or through temperature stabilization of the integrating sphere when high accuracy (<5%) measurements of flux, irradiance, or radiance are required from PTFE-based integrating spheres at temperatures close to the phase transition at 19°C.
烧结聚四氟乙烯(PTFE)具有高反射性,在遥感、辐射测量和光谱学中被广泛用作参考标准。在633nm单色波长下,已测得PTFE积分球在19°C室温相变过程中的输出通量相对变化为1.82±0.21%。输出通量的变化归因于PTFE总半球反射率发生了0.09±0.02%的微小变化,这是由相变导致其材料密度变化引起的。对于大多数用户而言,这种在总半球反射率中测得的微小变化不太可能对用作参考标准的PTFE平板反射器的精度产生显著影响。然而,由于积分球腔内会发生多次反射,这种影响会被放大,当需要在接近19°C相变温度下对基于PTFE的积分球进行高精度(<5%)的通量、辐照度或辐射率测量时,应通过数学校正或对积分球进行温度稳定化处理来采取补救措施。