一种用于测定人体中[75Se]硒代半胱氨酸保留量的共轭计数法。

A conjugate counting method to determine [75Se]SeHCAT retention in the human body.

作者信息

du Toit M D, Strydom W J, van Reenen O R, van der Merwe C F

机构信息

Department of Medical Physics, MEDUNSA, Republic of South Africa.

出版信息

Int J Rad Appl Instrum B. 1990;17(4):413-8. doi: 10.1016/0883-2897(90)90110-m.

Abstract

To evaluate the functional integrity of the distal part of the ileum the retention of a gamma-labelled bile acid (SeHCAT) in the human body can be measured with a detector. Due to the lack of a whole body counter at our institution a two detector system was designed to measure SeHCAT retention and an evaluation of such a system has been made. The detectors are positioned on either side of a patient lying supine on a hospital trolley. The trolley is stepped forward in 100 mm steps, to determine the SeHCAT activity in the patient. With these counts the location of the SeHCAT activity and total activity present in the body can be determined. A water filled phantom and a phantom consisting of nine 1-L saline bags with 75Se activity placed in them was used to determine system performance. Four patients with no history of bowel disease were compared with published data for normals. Results showed that the system performed satisfactorily, and accurate quantitative measurements could be made, showing that this inexpensive system could be used where a whole body counter is not available.

摘要

为评估回肠远端的功能完整性,可使用探测器测量人体中γ标记胆汁酸(硒-同型胆酸,SeHCAT)的潴留情况。由于我们机构没有全身计数器,因此设计了一种双探测器系统来测量SeHCAT潴留情况,并对该系统进行了评估。探测器放置在仰卧于医院推车上的患者身体两侧。推车以100毫米的步长向前移动,以确定患者体内的SeHCAT活性。通过这些计数,可以确定SeHCAT活性的位置以及体内存在的总活性。使用一个充水模体和一个由九个装有75Se活性的1升盐水袋组成的模体来确定系统性能。将四名无肠道疾病史的患者与正常人群的已发表数据进行比较。结果表明,该系统运行良好,可以进行准确的定量测量,这表明在没有全身计数器的情况下可以使用这种成本低廉的系统。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验