du Toit M D, Strydom W J, van Reenen O R, van der Merwe C F
Department of Medical Physics, MEDUNSA, Republic of South Africa.
Int J Rad Appl Instrum B. 1990;17(4):413-8. doi: 10.1016/0883-2897(90)90110-m.
To evaluate the functional integrity of the distal part of the ileum the retention of a gamma-labelled bile acid (SeHCAT) in the human body can be measured with a detector. Due to the lack of a whole body counter at our institution a two detector system was designed to measure SeHCAT retention and an evaluation of such a system has been made. The detectors are positioned on either side of a patient lying supine on a hospital trolley. The trolley is stepped forward in 100 mm steps, to determine the SeHCAT activity in the patient. With these counts the location of the SeHCAT activity and total activity present in the body can be determined. A water filled phantom and a phantom consisting of nine 1-L saline bags with 75Se activity placed in them was used to determine system performance. Four patients with no history of bowel disease were compared with published data for normals. Results showed that the system performed satisfactorily, and accurate quantitative measurements could be made, showing that this inexpensive system could be used where a whole body counter is not available.
为评估回肠远端的功能完整性,可使用探测器测量人体中γ标记胆汁酸(硒-同型胆酸,SeHCAT)的潴留情况。由于我们机构没有全身计数器,因此设计了一种双探测器系统来测量SeHCAT潴留情况,并对该系统进行了评估。探测器放置在仰卧于医院推车上的患者身体两侧。推车以100毫米的步长向前移动,以确定患者体内的SeHCAT活性。通过这些计数,可以确定SeHCAT活性的位置以及体内存在的总活性。使用一个充水模体和一个由九个装有75Se活性的1升盐水袋组成的模体来确定系统性能。将四名无肠道疾病史的患者与正常人群的已发表数据进行比较。结果表明,该系统运行良好,可以进行准确的定量测量,这表明在没有全身计数器的情况下可以使用这种成本低廉的系统。