Suppr超能文献

[空间外差光谱仪相位校正方法研究]

[Study on phase correction method of spatial heterodyne spectrometer].

作者信息

Wang Xin-Qiang, Ye Song, Zhang Li-Juan, Xiong Wei

机构信息

Department of Electronic Engineering and Automation, Guilin University of Electronic Technology, Guilin 541004, China.

出版信息

Guang Pu Xue Yu Guang Pu Fen Xi. 2013 May;33(5):1424-8.

Abstract

Phase distortion exists in collected interferogram because of a variety of measure reasons when spatial heterodyne spectrometers are used in practice. So an improved phase correction method is presented. The phase curve of interferogram was obtained through Fourier inverse transform to extract single side transform spectrum, based on which, the phase distortions were attained by fitting phase slope, so were the phase correction functions, and the convolution was processed between transform spectrum and phase correction function to implement spectrum phase correction. The method was applied to phase correction of actually measured monochromatic spectrum and emulational water vapor spectrum. Experimental results show that the low-frequency false signals in monochromatic spectrum fringe would be eliminated effectively to increase the periodicity and the symmetry of interferogram, in addition when the continuous spectrum imposed phase error was corrected, the standard deviation between it and the original spectrum would be reduced form 0.47 to 0.20, and thus the accuracy of spectrum could be improved.

摘要

在实际应用空间外差光谱仪时,由于多种测量原因,采集到的干涉图中存在相位畸变。因此,提出了一种改进的相位校正方法。通过傅里叶逆变换获得干涉图的相位曲线以提取单边变换光谱,在此基础上,通过拟合相位斜率得到相位畸变,进而得到相位校正函数,并对变换光谱与相位校正函数进行卷积处理以实现光谱相位校正。该方法应用于实际测量的单色光谱和模拟水汽光谱的相位校正。实验结果表明,能有效消除单色光谱条纹中的低频虚假信号,提高干涉图的周期性和对称性,此外,在对连续光谱施加相位误差进行校正时,其与原始光谱之间的标准偏差从0.47降低到0.20,从而提高了光谱的精度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验