Suppr超能文献

应用 ThinPrep 成像系统进行尿液细胞学检查:一项前瞻性研究。

Application of the ThinPrep imaging system in urine cytology: a prospective study.

机构信息

Department of Pathology, University Medical Center Groningen, Groningen, the Netherlands.

出版信息

Cancer Cytopathol. 2013 Aug;121(8):410-4. doi: 10.1002/cncy.21301. Epub 2013 Apr 23.

Abstract

BACKGROUND

In this prospective study, for the first time, the authors compared the accuracy of reading urine specimens using the ThinPrep Imager System (TIS) with the accuracy of conventional screening for the detection of abnormal urine cells.

METHODS

ThinPrep slides were made from 1455 urine specimens and were read with conventional screening and with TIS. Findings were categorized as "unsatisfactory or failure to read the slide," "benign," or "abnormal." The Cohen κ coefficient was calculated to determine inter-rater agreement between both methods. Urine samples that were followed by biopsies were used to compare the sensitivity, specificity, positive predictive value, and negative predictive value of both methods. From 22 urine specimens, the screening times were measured and compared.

RESULTS

There was substantial agreement between both methods (κ score, 0.77). Of 175 urine specimens that were followed by bladder biopsies, for conventional screening, the sensitivity was 51.3%, specificity was 68.4%, the positive predictive value was 77.2%, and the negative predictive value was 40.2%; for TIS screening, the respective values were 54.6%, 68.4%, 78.3%, and 41.9%. The average time was 5.2 minutes for conventional screening and 3.9 minutes for TIS.

CONCLUSIONS

With a κ score of 0.77, the current study demonstrated a good correlation between reading urine specimens with conventional screening and with TIS. Using TIS resulted in slightly increased sensitivity, positive predictive value, and negative predictive value compared with conventional screening and had the same specificity. These results indicate that reading urine specimens using TIS is equally reliable as conventional cytology.

摘要

背景

在这项前瞻性研究中,作者首次比较了使用 ThinPrep Imager System(TIS)读取尿液标本的准确性与传统筛查方法检测异常尿液细胞的准确性。

方法

从 1455 份尿液标本中制作 ThinPrep 载玻片,并使用传统筛查和 TIS 进行读取。结果分为“不满意或无法读取载玻片”、“良性”或“异常”。计算 Cohen κ 系数以确定两种方法之间的观察者间一致性。使用活检随访的尿液样本比较两种方法的敏感性、特异性、阳性预测值和阴性预测值。从 22 份尿液标本中测量并比较了筛查时间。

结果

两种方法之间存在高度一致性(κ 评分,0.77)。在 175 份随后进行膀胱活检的尿液标本中,对于传统筛查,敏感性为 51.3%,特异性为 68.4%,阳性预测值为 77.2%,阴性预测值为 40.2%;对于 TIS 筛查,相应值分别为 54.6%、68.4%、78.3%和 41.9%。传统筛查的平均时间为 5.2 分钟,TIS 为 3.9 分钟。

结论

当前研究的 κ 评分为 0.77,表明使用传统筛查和 TIS 读取尿液标本之间存在良好的相关性。与传统筛查相比,TIS 略提高了敏感性、阳性预测值和阴性预测值,特异性相同。这些结果表明,使用 TIS 读取尿液标本与传统细胞学一样可靠。