Cholewicki J, McGill S M, Wells R P, Vernon H
Occupational Biomechanics Laboratory, Department of Kinesiology, University of Waterloo, Ontario, Canada.
Clin Biomech (Bristol). 1991 May;6(2):73-8. doi: 10.1016/0268-0033(91)90002-8.
Optical distortions and digitizing errors must be minimized to obtain accurate measurements of spine kinematics from videofluoroscopic imaging. This study adapted and evaluated a method for reducing these errors. Image and signal processing techniques were applied to obtain lumbar spine kinematic measures from walking and trunk extension trials. Mean absolute error was conservatively estimated to be 0.69° (sd = 0.43°) for vertebral rotation, independent of angle, and 0.33 mm (sd = 0.25 mm) for linear measurements. Digital filtering was applied to further minimize random errors.
为了从视频荧光成像中获得准确的脊柱运动学测量结果,必须将光学畸变和数字化误差降至最低。本研究采用并评估了一种减少这些误差的方法。应用图像和信号处理技术从步行和躯干伸展试验中获取腰椎运动学测量值。椎体旋转的平均绝对误差保守估计为0.69°(标准差=0.43°),与角度无关,线性测量的平均绝对误差为0.33毫米(标准差=0.25毫米)。应用数字滤波进一步减少随机误差。