Suppr超能文献

Simple device for air reduction of intussusception.

作者信息

Shiels W E, Bisset G S, Kirks D R

机构信息

Department of Radiology, University of Cincinnati College of Medicine, Ohio.

出版信息

Pediatr Radiol. 1990;20(6):472-4. doi: 10.1007/BF02075213.

Abstract

A new device has been developed for pneumatic reduction of childhood intussusception. The device consists of a reusable, hand-held pressure gauge and insufflator connected to a disposable enema tip and tubing system. The assembled device permits one-handed operation which maximizes operator control of the air reduction procedure. The system has been used to evaluate 50 cases of suspected intussusception. Reduction was successful in 19 of 22 confirmed intussusceptions. This device provides a practical, simple and safe method for diagnosis and treatment of intussusception.