Chemical Resource Beneficiation, Membrane Technology Group, North-West University, Private Bag X6001, Potchefstroom 2520, South Africa.
J Chromatogr A. 2013 Sep 13;1307:180-90. doi: 10.1016/j.chroma.2013.07.101. Epub 2013 Aug 1.
A dual-channel gas chromatographic method is described in this paper that can be conveniently used for quantitation of NF3/CF4 mixtures with a thermal conductivity detector (TCD) on one channel for the quantitation of high-concentrations, and a pulsed discharge helium ionization detector (PDHID) on a second channel for the quantitation of low concentrations. It is shown that adequate separation is achieved on both channels with this dual single-column setup in which column switching as used for NF3/CF4 analysis in industrial chromatographic methods are not required, thus yielding an effective analysis method for laboratory-scale investigations. In addition, the use of packed columns with purified divinylbenzene-styrene co-polymers as the sole stationary phase yields satisfactory resolution between NF3 and CF4 at isothermal conditions of 30°C, with elution times of less than 8min on the TCD channel and less than 4min on the PDHID channel. Consequently, this method allows for reliable, straight-forward quantitation of NF3/CF4 mixtures, which is necessary when studying the commercially important problem of NF3 and CF4 separation by different methods. Therefore, the applicability of the method to studying membrane separation of NF3 and CF4 is briefly discussed and illustrated, for which the dual-channel setup is especially beneficial.
本文介绍了一种双通道气相色谱法,该方法可方便地用于定量分析 NF3/CF4 混合物。其中一个通道采用热导检测器(TCD)定量分析高浓度,另一个通道采用脉冲放电氦离子化检测器(PDHID)定量分析低浓度。结果表明,在双通道单柱装置中,通过这种双柱单柱装置可以实现充分的分离,无需进行工业色谱方法中用于 NF3/CF4 分析的柱切换,从而为实验室规模的研究提供了有效的分析方法。此外,在 30°C 的等温条件下,使用纯化的二乙烯基苯-苯乙烯共聚物作为唯一固定相的填充柱,可在 TCD 通道上的洗脱时间小于 8min,在 PDHID 通道上的洗脱时间小于 4min,从而在 NF3 和 CF4 之间获得令人满意的分辨率。因此,当研究通过不同方法分离 NF3 和 CF4 这一具有商业重要性的问题时,该方法可用于可靠、直接地定量分析 NF3/CF4 混合物。因此,本文简要讨论并说明了该方法在研究 NF3 和 CF4 的膜分离中的适用性,双通道设置在这方面特别有益。