Nuffield Department of Surgical Sciences, Oxford University Hospitals, Oxford, UK.
Ultrasound Med Biol. 2013 Oct;39(10):1785-93. doi: 10.1016/j.ultrasmedbio.2013.04.010. Epub 2013 Aug 9.
High-intensity focused ultrasound (HIFU) is an attractive therapy for kidney cancer, but its efficacy can be limited by heat deposition in the pre-focal tissues, notably in fat around the kidney (peri-nephric fat), the acoustic properties of which have not been well characterized. Measurements of attenuation were made using a modified insertion-loss technique on fresh, unfixed peri-nephric fat obtained from patients undergoing kidney surgery for cancer. The de-focusing effect of changing the position of the fat layers was also investigated using fresh subcutaneous fat from euthanized pigs. The mean attenuation of human peri-nephric fat was found to be 11.9 ± 0.9 Np/m (n = 10) at 0.8 MHz, the frequency typically used for HIFU ablation of kidney tumors, with a frequency dependence of f(1.2). A typical 2- to 4-cm thickness of peri-nephric fat would result in a de-rated intensity of 3% - 62% at 0.8 MHz compared with a hypothetical patient with no peri-nephric fat. Through the use of freshly excised porcine subcutaneous fat, the presence of fat 100 mm in front of the focus was found to have a de-focusing effect of approximately 1 mm in both transverse directions, which corresponds to a full HIFU beam width off-target. Peri-nephric fat may significantly affect both the intensity and accuracy of HIFU fields used for the ablation of kidney cancer.
高强度聚焦超声(HIFU)是治疗肾癌的一种有吸引力的疗法,但由于焦点前组织(特别是肾周围脂肪)中的热量沉积,其疗效可能受到限制,而肾周围脂肪的声学特性尚未得到很好的描述。在对接受肾癌手术的患者的新鲜、未固定的肾周脂肪进行的一项修改后的插入损耗技术测量中,对衰减进行了测量。还使用从安乐死猪身上获取的新鲜皮下脂肪研究了改变脂肪层位置的散焦效应。在 0.8MHz(通常用于 HIFU 消融肾肿瘤的频率)下,发现人肾周脂肪的平均衰减率为 11.9±0.9Np/m(n=10),频率依赖性为 f(1.2)。典型的 2-4cm 厚的肾周脂肪会导致 0.8MHz 时的强度降低 3%-62%,与没有肾周脂肪的假设患者相比。通过使用新鲜切除的猪皮下脂肪,发现焦点前 100mm 处的脂肪在横向上有大约 1mm 的散焦效应,这相当于全 HIFU 光束的脱靶。肾周脂肪可能会显著影响用于消融肾癌的 HIFU 场的强度和准确性。