应用眼前节光学相干断层扫描仪观察飞秒激光小切口角膜基质透镜取出术(SMILE)中角膜瓣的形态及其与屈光效果的关系。

The morphology of corneal cap and its relation to refractive outcomes in femtosecond laser small incision lenticule extraction (SMILE) with anterior segment optical coherence tomography observation.

机构信息

Department of Ophthalmology, Eye and ENT Hospital of Fudan University, Myopia Key Laboratory of the Health Ministry, Shanghai, China.

出版信息

PLoS One. 2013 Aug 5;8(8):e70208. doi: 10.1371/journal.pone.0070208. Print 2013.

Abstract

PURPOSE

To investigate the morphology of corneal caps in femtosecond laser small incision lenticule extraction (SMILE) and its relation to the refractive outcomes.

METHODS

A prospective study of fifty-four corneal caps created with VisuMax femtosecond laser were examined using an Fourier-domain optical coherence tomography at 1 day, 1 week, 1 month and 6 months after SMILE. The cap thickness at nine points on each of the four meridians (0°, 45°, 90°, 135°) and the diameter were measured. Cap morphology, changes over time and its correlation with refractive outcomes were assessed.

RESULTS

The mean achieved central cap thickness were (108.74±5.06) µm at 6 months and (107.32±4.81 ) µm at 1 month postoperatively, significantly thinner than that at 1 day (110.81±7.95) µm and 1 week (109.58±7.48 ) µm (P<0.05). The mean diameter on 0° meridian was (7.61±0.07) mm, significantly larger than that on 90° meridian (7.57±0.06) mm (P = 0.001). Cap morphology showed good regularity, except that the differences of points in two pairs were significant at 1 day postoperatively. The uniformity was consistent over time and the central cap thickness was thinner than those in the paracentral and peripheral areas. The refractive outcomes stabilized within 1 month. Uncorrected distance visual acuity (UDVA) was correlated to the central cap thickness at 1 day and 1 week (both rs  = 0.33, p<0.05). The uniformity index was correlated with UDVA (rs  = 0.34, p<0.05) and corrected distance visual acuity (rs  = 0.32, p<0.05) at 1 week postoperatively.

CONCLUSIONS

Corneal caps of SMILE are predictable with good reproducibility, regularity and uniformity. Cap morphology might have a mild effect on refractive outcomes in the early stage. Further study should focus on the impact on the visual quality.

摘要

目的

研究飞秒激光小切口角膜微透镜取出术(SMILE)中角膜帽的形态及其与屈光结果的关系。

方法

前瞻性研究 54 例使用 VisuMax 飞秒激光制作的角膜帽,分别于 SMILE 术后 1 天、1 周、1 个月和 6 个月,采用傅里叶域光学相干断层扫描(OCT)进行检查。在 4 条子午线上的 9 个点测量角膜帽的厚度(0°、45°、90°、135°)和直径。评估角膜帽形态、随时间的变化及其与屈光结果的相关性。

结果

术后 6 个月中央角膜帽的平均实际厚度为(108.74±5.06)μm,术后 1 个月为(107.32±4.81)μm,明显低于术后 1 天(110.81±7.95)μm 和术后 1 周(109.58±7.48)μm(P<0.05)。0°子午线上的平均直径为(7.61±0.07)mm,明显大于 90°子午线上的直径(7.57±0.06)mm(P = 0.001)。角膜帽形态具有较好的规律性,除术后 1 天 4 对测量点之间存在差异有统计学意义外。角膜帽形态的均匀性随时间推移而保持一致,中央角膜帽厚度较周边和旁中央区薄。屈光结果在 1 个月内稳定。术后 1 天和 1 周的未矫正远视力(UDVA)与中央角膜帽厚度呈正相关(rs  = 0.33,p<0.05)。术后 1 周的均匀指数与 UDVA(rs  = 0.34,p<0.05)和矫正远视力(rs  = 0.32,p<0.05)呈正相关。

结论

SMILE 术后角膜帽具有可预测性、良好的重现性、规律性和均匀性。在早期阶段,角膜帽形态可能对屈光结果有轻微影响。进一步的研究应侧重于对视觉质量的影响。