1 Department of Health Science and Technology, Aalborg University , Aalborg, Denmark .
Diabetes Technol Ther. 2013 Oct;15(10):825-35. doi: 10.1089/dia.2013.0041. Epub 2013 Aug 14.
The deviation of continuous subcutaneous glucose monitoring (CGM) data from reference blood glucose measurements is substantial, and adequate signal processing is required to reduce the discrepancy between subcutaneous glucose and blood glucose values. The purpose of this study was to develop a multistep algorithm for the processing and calibration of continuous subcutaneous glucose monitoring data with high accuracy and short delay. Algorithm
The algorithm comprises three steps: rate-limiting filtering, selective smoothing, and robust calibration. Initially, the algorithm detects nonphysiological glucose rate-of-change and corrects it with a weighted local polynomial. Noisy signal parts that require smoothing are then detected based on zero crossing count of the sensor signal first-order differences, and an exponentially weighted moving average smooths the noisy parts of the signal afterward. Finally, calibration is performed using a first-order polynomial as the conversion function, with coefficients being estimated using robust regression with a bi-square weight function. ALGORITHM PERFORMANCE: The performance of the algorithm was evaluated on 16 patients with type 1 diabetes mellitus. To compare the algorithm with state-of-the-art CGM data denoising and calibration, the rate-limiting filter and selective smoothing were replaced with an adaptive Kalman filter, and the calibration method was replaced with the calibration algorithm presented in one of the Medtronic (Northridge, CA) CGM patents. The median (mean) of the absolute relative deviation (ARD) of the sensor glucose values processed by the newly developed algorithm from capillary reference blood glucose measurements was 14.8% (22.6%), 10.6% (14.6%), and 8.9% (11.7%) in hypoglycemia, euglycemia, and hyperglycemia, respectively, whereas for the alternative algorithm, the median (mean) was 22.2% (26.9%), 12.1% (15.9%), and 8.8 (11.3%), respectively. The median (mean) ARD in all ranges was 10.3% (14.7%) for the new algorithm and 11.5% (15.8%) for the alternative algorithm. The new algorithm had an average delay of 2.1 min across the patients, and the alternative algorithm had an average delay of 2.9 min.
The presented algorithm may increase the accuracy of CGM data.
连续皮下葡萄糖监测(CGM)数据与参考血糖测量值存在较大偏差,需要进行适当的信号处理,以减少皮下葡萄糖与血糖值之间的差异。本研究旨在开发一种高精度、低延迟的多步骤算法,用于处理和校准连续皮下葡萄糖监测数据。
该算法由三个步骤组成:限速滤波、选择性平滑和稳健校准。首先,算法检测非生理葡萄糖变化率,并使用加权局部多项式进行校正。然后,根据传感器信号一阶差分的过零点计数检测需要平滑的噪声信号部分,并使用指数加权移动平均对信号的噪声部分进行平滑处理。最后,使用一阶多项式作为转换函数进行校准,使用双平方权重函数的稳健回归估计系数。
该算法在 16 例 1 型糖尿病患者中进行了评估。为了将该算法与最先进的 CGM 数据去噪和校准进行比较,限速滤波器和选择性平滑被自适应卡尔曼滤波器取代,而校准方法则被美敦力(加利福尼亚州北岭)CGM 专利中提出的校准算法所取代。新开发算法处理后的传感器葡萄糖值与毛细血管参考血糖测量值的平均绝对偏差(ARD)中位数(平均值)分别为低血糖、血糖正常和高血糖时的 14.8%(22.6%)、10.6%(14.6%)和 8.9%(11.7%),而替代算法的中位数(平均值)分别为 22.2%(26.9%)、12.1%(15.9%)和 8.8(11.3%)。在所有范围内,新算法的中位数(平均值)为 10.3%(14.7%),替代算法的中位数(平均值)为 11.5%(15.8%)。新算法的平均延迟为 2.1 分钟,替代算法的平均延迟为 2.9 分钟。
所提出的算法可以提高 CGM 数据的准确性。