In-vitro assessment of centrifugal pumps for ventricular assist.

作者信息

Jakob H, Kutschera Y, Palzer B, Prellwitz W, Oelert H

机构信息

Department of Cardiothoracic and Vascular Surgery, Mainz University Hospital, F.R.G.

出版信息

Artif Organs. 1990 Aug;14(4):278-83. doi: 10.1111/j.1525-1594.1990.tb02969.x.

Abstract

Currently two major types of centrifugal pumps are commercially available for ventricular assist: the Biomedicus-cone (Group I) and the Centrimed-impeller pump (now Sarns 3M) (Group II). To compare them for blood trauma and hemolysis, an in-vitro experiment was designed with a Stöckert roller pump as a standard control (Group III). The in-vitro circuit was constructed consisting of a pump head, electromagnetic flow probe, polyvinyl chloride tubing and a reservoir, identical for all groups. Human ACD blood was used for priming and was circulated with a flow rate of 2 L/min for 24 h. Blood samples were taken at 0, 1, 3, 6, 12, and 24 h and zero control values were subtracted from the resulting data per time interval. Among the 16 parameters studied, a highly significant difference in favor of Group I was found for glutamate oxalacetate transaminase (GOT) and lactate dehydrogenase (p less than 0.0001) and for the free plasma hemoglobin (p less than 0.0001) after 12 and 24 h, respectively. The hemolytic index (Allen) again was lowest for group I in contrast to Groups II and III (0.012 versus 0.060 and 1.70) after 24 h. All other parameters studied did not render significant differences between the systems tested. The authors conclude that the Biomedicus pump currently is the least traumatic centrifugal pump for ventricular assist.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验