Department of Ophthalmology, Faculty of Medicine, University of Tsukuba, Ibaraki, Japan.
Am J Ophthalmol. 2013 Nov;156(5):941-947.e1. doi: 10.1016/j.ajo.2013.06.016. Epub 2013 Aug 20.
To investigate intraocular pressure (IOP) fluctuation during various vitrectomy maneuvers using the vitrectomy system (Alcon Constellation Vision System).
An experimental study as laboratory investigation.
In porcine eyes, 23- and 25-gauge vitrectomy was performed, and IOP fluctuations were evaluated in vitreous cutting mode, in aspiration mode, and during scleral compression. The measurements were performed with the IOP control setting turned on or off.
Using the 23-gauge system with the IOP control setting turned on, IOP decreased from 30 to 23.7 mm Hg after starting vitreous cutting, and then returned to 30 mm Hg in 2.6 seconds. When the IOP control setting was turned off, IOP decreased to 19.1 mm Hg in 0.9 seconds, and remained at that pressure. Under aspiration at 650 mm Hg without cutting, IOP showed a sharp depression from 30 to 12.2 mm Hg, and then returned to 30.6 mm Hg in 2.6 seconds with the IOP control setting turned on. When the IOP control setting was turned off, IOP decreased to 2.2 mm Hg in 9.7 seconds, and did not recover. When the sclera was compressed without aspiration, IOP rapidly increased to 70-100 mm Hg, and then slowly decreased to 30 mm Hg in 3.5-4.0 seconds, with or without the IOP control system. Similar data were obtained with 25-gauge vitrectomy.
The IOP control system can attenuate IOP fluctuations during vitrectomy maneuvers. There was no significant difference in IOP fluctuations between 23- and 25-gauge systems.
使用玻璃体切割系统(爱尔康康柏视觉系统)研究各种玻璃体切割操作过程中的眼压(IOP)波动。
实验性研究,实验室研究。
在猪眼上进行 23 号和 25 号玻璃体切割,评估玻璃体切割模式、抽吸模式和巩膜压缩时的 IOP 波动。测量时开启或关闭眼压控制设置。
使用带眼压控制设置的 23 号系统,开始玻璃体切割后,IOP 从 30 降至 23.7mmHg,然后在 2.6 秒内恢复至 30mmHg。当眼压控制设置关闭时,IOP 在 0.9 秒内降至 19.1mmHg,并保持在该压力下。在不切割的情况下以 650mmHg 进行抽吸时,IOP 从 30 急剧降至 12.2mmHg,然后在开启眼压控制设置的情况下,2.6 秒内恢复至 30.6mmHg。当眼压控制设置关闭时,IOP 在 9.7 秒内降至 2.2mmHg,且无法恢复。在不抽吸的情况下压缩巩膜时,IOP 迅速升高至 70-100mmHg,然后在 3.5-4.0 秒内缓慢降至 30mmHg,开启或不开启眼压控制系统均如此。25 号玻璃体切割也获得了类似的数据。
眼压控制系统可减轻玻璃体切割操作过程中的眼压波动。23 号和 25 号系统的眼压波动无显著差异。