加速聚乙烯磨损和非骨水泥 Reflection 髋臼组件早期失效:十年至十五年随访研究。

Accelerated polyethylene wear and early failure of the uncemented Reflection acetabular component: a ten to fifteen-year follow-up study.

机构信息

Department of Orthopaedics and Trauma, The Royal Infirmary of Edinburgh, Little France, Edinburgh, EH16 4SA, UK,

出版信息

Int Orthop. 2013 Dec;37(12):2337-43. doi: 10.1007/s00264-013-2058-3. Epub 2013 Aug 22.

Abstract

PURPOSE

The Reflection® second-generation uncemented acetabular component was designed to address increased rates of failure observed with the early acetabular designs. However, the reported survivorship of this acetabular component has been conflicting. The aim of this study was to describe the ten to 15-year survivorship and polyethylene wear rate for the uncemented Reflection® acetabular component performed as part of a primary total hip replacement.

METHODS

One hundred and four consecutive Reflection® uncemented acetabular components in 97 patients were identified from a prospective arthroplasty database with a minimum of ten years of follow-up. No patient was lost to follow-up. Mean cohort age was 59.1 years.

RESULTS

There were 24 revisions of the acetabular component. The all-cause survival rate at ten years was 77.2% [95% confidence interval (CI) 73.9-80.5]. Mean linear wear was 0.20 mm [standard deviation (SD) 0.08] per year and the volumetric wear was 106.2 mm(3) per year. At final follow-up, more than half the patients had osteolysis identified around the femoral component. The mean Oxford Hip Score was 35.6 (SD 9.8) at a mean follow-up of 11.9 years, with six (11.1%) excellent, 26 (48.1%) good, 17 (31.5%) fair and five (9.3%) poor outcomes. Despite the fact that more than a third had a fair or poor outcome, only four (7.4%) were not satisfied with their hip.

CONCLUSIONS

Due to the high rate of relatively asymptomatic polyethylene wear and osteolysis associated with this acetabular component, in our department, we now review all surviving patients both clinically and radiographically on an annual basis.

摘要

目的

Reflection® 第二代非骨水泥髋臼组件的设计旨在解决早期髋臼设计中观察到的失败率增加的问题。然而,该髋臼组件的报告存活率存在冲突。本研究的目的是描述作为初次全髋关节置换术一部分的非骨水泥 Reflection®髋臼组件的十年至十五年存活率和聚乙烯磨损率。

方法

从前瞻性关节置换数据库中确定了 97 名患者的 104 个连续的 Reflection® 非骨水泥髋臼组件,随访时间至少为十年。没有患者失访。平均队列年龄为 59.1 岁。

结果

共有 24 例髋臼组件翻修。十年时的全因存活率为 77.2%[95%置信区间(CI)73.9-80.5]。平均线性磨损率为每年 0.20 毫米[标准差(SD)0.08],体积磨损率为每年 106.2 毫米(3)。在最终随访时,超过一半的患者在股骨组件周围发现了骨溶解。平均牛津髋关节评分在 11.9 年的平均随访时间为 35.6(SD 9.8),其中 6 例(11.1%)为优秀,26 例(48.1%)为良好,17 例(31.5%)为一般,5 例(9.3%)为差。尽管超过三分之一的患者有一般或较差的结果,但只有 4 例(7.4%)对他们的髋关节不满意。

结论

由于该髋臼组件与相对无症状的聚乙烯磨损和骨溶解率较高,在我们部门,我们现在每年对所有存活患者进行临床和影像学检查。