Department of Radiology, Aarhus University Hospital, Aarhus, Denmark.
Br J Radiol. 2013 Nov;86(1031):20130257. doi: 10.1259/bjr.20130257. Epub 2013 Sep 12.
To examine whether dynamic contrast-enhanced CT (DCE-CT) could be used to characterise and safely distinguish between malignant and benign lung tumours in patients with suspected lung cancer.
Using a quantitative approach to DCE-CT, two separate sets of regions of interest (ROIs) in tissues were placed in each tumour: large ROIs over the entire tumour and small ROIs over the maximally perfused parts of the tumour. Using mathematical modelling techniques and dedicated perfusion software, this yielded a plethora of results.
First, because of their non-normal distribution, DCE-CT measurements must be analysed using log scale data transformation. Second, there were highly significant differences between large ROI and small ROI measurements (p<0.001). Thus, the ROI method used in a given study should always be specified in advance. Third, neither quantitative parameters (blood flow and blood volume) nor semi-quantitative parameters (peak enhancement) could be used to distinguish between malignant and benign tumours. This was irrespective of the method of quantification used for large ROIs (0.13<p<0.76) and small ROIs (0.084<p<0.31). Fourth, although there were no indications of systematic reproducibility bias, the 95% limits of agreement were so broad that the risk of disagreement between the measurements could affect the clinical use of the measurements. This lack of reproducibility should be addressed. CONCLUSION AND ADVANCES IN KNOWLEDGE: A quantitative approach to DCE-CT is not a clinically usable method for characterising lung tumours.
研究动态对比增强 CT(DCE-CT)是否可用于对疑似肺癌患者的良、恶性肺部肿瘤进行特征描述和安全区分。
使用 DCE-CT 的定量方法,在每个肿瘤中分别在组织中放置两个独立的感兴趣区(ROI):大 ROI 覆盖整个肿瘤,小 ROI 覆盖肿瘤的最大灌注部分。使用数学建模技术和专用灌注软件,得到了大量的结果。
首先,由于 DCE-CT 测量值呈非正态分布,因此必须使用对数标度数据转换进行分析。其次,大 ROI 和小 ROI 测量值之间存在高度显著差异(p<0.001)。因此,在给定的研究中,应该事先指定 ROI 方法。第三,无论是定量参数(血流量和血容量)还是半定量参数(峰值增强)都无法用于区分良、恶性肿瘤。这与大 ROI(0.13<p<0.76)和小 ROI(0.084<p<0.31)使用的定量方法无关。第四,尽管没有系统可重复性偏差的迹象,但 95%的一致性界限如此之宽,以至于测量值之间的不一致性可能会影响测量值的临床应用。这种缺乏可重复性应该得到解决。
DCE-CT 的定量方法不是一种可用于描述肺部肿瘤的临床方法。