Respir Physiol Neurobiol. 2013 Dec 1;189(3):473-6. doi: 10.1016/j.resp.2013.08.023. Epub 2013 Sep 11.
Optoelectronic plethysmography (OEP) has been used to measure changes in chest wall volume and its compartments. However, literature lacks research on its reliability. The purpose of this study was to evaluate the intra-rater and inter-rater reliability of OEP. Thirty-two healthy subjects were evaluated at rest and during submaximal exercise on a cycle ergometer. The following variables were assessed: chest wall volume (VCW); percentage contribution of the pulmonary rib cage (V(rcp)%), abdominal rib cage (V(rca)%), rib cage (V(rc)%) and abdomen (Vab%); chest wall end-expiratory volume (Vee(cw)); chest wall end-inspiratory volume (Vei(cw)); ratio of inspiratory time to total time of the respiratory cycle (Ti/T(tot)); respiratory rate (f) and mean inspiratory flow (V(cw)/Ti). Intraclass correlation coefficient (ICC) and coefficient of variation of Method Error (CV(ME)) were used to evaluate reliability. Results showed ICC values higher than 0.75 and CV(ME) values less than 10% for most variables at rest and during exercise indicating that OEP is a reliable instrument to assess chest wall volumes at rest and during exercise in healthy subjects.
光电容积描记法(OEP)已被用于测量胸壁容积及其各腔室的变化。然而,文献中缺乏对其可靠性的研究。本研究旨在评估 OEP 的内部和外部可靠性。32 名健康受试者在休息时和在自行车测力计上进行亚最大运动时接受了评估。评估了以下变量:胸壁容积(VCW);肺肋骨笼(V(rcp)%)、腹部肋骨笼(V(rca)%)、肋骨笼(V(rc)%)和腹部(Vab)%的百分比贡献;胸壁呼气末容积(Vee(cw));胸壁吸气末容积(Vei(cw));呼吸周期的吸气时间与总时间之比(Ti/T(tot));呼吸频率(f)和平均吸气流量(V(cw)/Ti)。内部一致性系数(ICC)和方法误差变异系数(CV(ME))用于评估可靠性。结果表明,在休息和运动时,大多数变量的 ICC 值高于 0.75,CV(ME)值低于 10%,这表明 OEP 是一种可靠的仪器,可用于评估健康受试者在休息和运动时的胸壁容积。