ActiGraph™ 活动监测器:“固件效应”。

ActiGraph™ activity monitors: "the firmware effect".

机构信息

1Northeastern University, Boston, MA; and 2University of Massachusetts, Amherst, MA.

出版信息

Med Sci Sports Exerc. 2014 Apr;46(4):834-9. doi: 10.1249/MSS.0000000000000145.

Abstract

PURPOSE

The objective of this study is to examine the effect of different firmware versions on ActiGraph™ counts from the laboratory, field, and mechanical shaker testing.

METHODS

Counts from 5 GT3X and 7 GT1M firmware versions were compared in this study. Monitors uploaded with these firmware versions were worn on the hip by 15 participants (age = 24.9 ± 5.0 yr, BMI= 23.9 ± 2.4 kg · m(-2)) who performed laboratory-based treadmill (walking: 1.5, 3.0 and 4.5 mph; running: 6 mph) and simulated free-living activities (sitting, self-paced walking, filing papers, dusting, vacuuming, and cleaning the room). Testing was also conducted during 1 d of free living and using orbital mechanical shaker testing at 0.7, 1.3, 2.0, and 3.0 Hz. Intermonitor comparisons for vertical, anteroposterior, mediolateral, and triaxial vector magnitude counts were conducted using one-way ANOVA and post hoc pairwise comparisons (P < 0.05).

RESULTS

Vertical counts during treadmill walking at 1.5 mph from the GT1M monitor with firmware version 1.1.0. were significantly greater (P < 0.05; 75%) than output from the monitor with firmware version 1.3.0. Shaker testing revealed statistically significant differences in vertical and lateral counts. Although there were no significant differences among activity counts in the free-living comparisons, firmware version 1.1.0. produced the highest vertical counts during this protocol.

CONCLUSION

Greater sensitivity of firmware version 1.1.0. to low-frequency sedentary activities resulted in greater counts than other firmware versions. It is recommended that before releasing new firmware, ActiGraph™ perform both human and mechanical shaker testing to verify comparability in outputs between new and previous firmware versions.

摘要

目的

本研究旨在检验不同固件版本对实验室、现场和机械振动机测试中 ActiGraph™ 计数值的影响。

方法

本研究比较了 5 种 GT3X 和 7 种 GT1M 固件版本的计数值。将这些固件版本上传到监测器后,让 15 名参与者(年龄=24.9±5.0 岁,BMI=23.9±2.4kg·m(-2))佩戴在髋部,进行基于实验室的跑步机(步行:1.5、3.0 和 4.5mph;跑步:6mph)和模拟自由生活活动(坐、自主步行、整理文件、除尘、吸尘和打扫房间)。测试还在 1 天的自由生活中进行,并在 0.7、1.3、2.0 和 3.0Hz 的轨道机械振动机测试中进行。使用单向方差分析和事后成对比较(P<0.05)对垂直、前-后、内-外侧和三轴矢量幅度计数值进行了监测器间比较。

结果

GT1M 监测器的固件版本 1.1.0 在 1.5mph 跑步机步行时的垂直计数明显高于固件版本 1.3.0 的输出(P<0.05;75%)。振动机测试显示垂直和水平计数存在统计学差异。虽然在自由生活比较中活动计数没有显著差异,但在该协议中,固件版本 1.1.0 产生了最高的垂直计数。

结论

固件版本 1.1.0 对低频静坐活动的更高敏感度导致计数高于其他固件版本。建议在发布新固件之前,ActiGraph™ 进行人体和机械振动机测试,以验证新固件和旧固件版本之间输出的可比性。