系统荧光透视和电评估植入硅胶-聚氨基甲酸乙酯共聚物(Optim™)涂层导联的植入式心脏复律除颤器患者。

Systematic fluoroscopic and electrical assessment of implantable cardioverter-defibrillator patients implanted with silicone-polyurethane copolymer (Optim™) coated leads.

机构信息

Department of Internal Medicine, Division of Cardiology, University Hospital of Tor Vergata, Viale Oxford, 81, 00133 Rome, Italy.

出版信息

Europace. 2014 Feb;16(2):265-70. doi: 10.1093/europace/eut236. Epub 2013 Sep 15.

Abstract

AIMS

Serious concerns have been recently raised about the reliability of the silicone-polyurethane copolymer (Optim™) lead insulation system. We sought to identify insulation defects and Optim-lead failures by systematic fluoroscopic and electrical assessment in a prospectively defined cohort of implantable cardioverter-defibrillator (ICD) patients.

METHODS AND RESULTS

Between July 2007 and December 2011, 234 patients were implanted with 413 optim-coated leads as part of an ICD system at a single centre. Fluoroscopic screening with high-resolution cine-fluoroscopy at 30 frames per second was offered to all patients. In addition, the electrical integrity of all implanted leads was assessed. Durata, Riata ST Optim, and low-voltage Optim leads were implanted in 199, 26, and 188 cases, respectively. During a total follow-up of 10 036 lead-months, there were 7 Optim-lead failures (defined as electrical malfunction resulting in lead replacement) and 31 deaths; no cases of electrical noises were encountered. The overall incidence of lead failure was 1.2 vs. 0.3 per 100 lead-years, for high- and low-voltage leads, respectively (P = 0.1). One hundred fifty-one patients agreed to undergo fluoroscopy screening; none of the 264 analysed Optim leads were found to have any fluoroscopically visible structural defects after an average of 31 months post-implant.

CONCLUSION

This study represents the first systematic screening of Optim-coated leads in a large unselected cohort of ICD patients. Over a 5-year period few lead failures were observed and normal fluoroscopic appearance was present in all patients.

摘要

目的

最近对硅-聚碳酸酯共聚物(Optim™)导联绝缘系统的可靠性提出了严重关切。我们通过对植入式心脏复律除颤器(ICD)患者的前瞻性队列进行系统的透视和电评估,试图确定绝缘缺陷和 Optim 导联故障。

方法和结果

在 2007 年 7 月至 2011 年 12 月期间,在一家单中心为 234 名患者植入了 413 个 Optim 涂层导联,作为 ICD 系统的一部分。为所有患者提供每秒 30 帧的高分辨率电影透视荧光检查。此外,还评估了所有植入导联的电完整性。植入了 Durata、Riata ST Optim 和低压 Optim 导联,分别有 199、26 和 188 例。在总共 10036 个导联月的随访期间,有 7 个 Optim 导联发生故障(定义为导致导联更换的电故障)和 31 例死亡;未发现电噪声病例。高、低压导联的导联总故障率分别为 1.2/100 导联年和 0.3/100 导联年(P = 0.1)。151 名患者同意接受透视筛查;在植入平均 31 个月后,对 264 个分析的 Optim 导联中没有一个在透视下发现任何结构缺陷。

结论

本研究代表了在大规模未选择的 ICD 患者队列中对 Optim 涂层导联的首次系统筛查。在 5 年期间,观察到的导联故障很少,所有患者的透视外观均正常。