van Wijk Diederik F, Strang Aart C, Duivenvoorden Raphael, Enklaar Dirk-Jan F, van der Geest Rob J, Kastelein John J P, de Groot Eric, Stroes Erik S G, Nederveen Aart J
Department of Vascular Medicine, Academic Medical Center, Room F4-159.2, PO Box 22660, 1100 DD, Amsterdam, The Netherlands,
MAGMA. 2014 Jun;27(3):219-26. doi: 10.1007/s10334-013-0407-2. Epub 2013 Sep 18.
To improve carotid 3T magnetic resonance imaging (MRI) dimension measurements in patients with overt atherosclerotic carotid artery disease.
In 31 patients with advanced atherosclerotic carotid artery disease, two high resolution (0.25 × 0.25 mm(2); HR) and two routinely used low resolution (0.50 × 0.50 mm(2); LR) carotid 3T MRI scans were performed within 1 month. After manual delineation of carotid wall contours in a dedicated image analyses program in eight slices covering the atherosclerotic plaque, image reproducibility, as well as the within-reader and between-reader variability were determined.
We found significantly higher intraclass correlation coefficients for total wall volume, mean wall area and mean wall thickness for the HR measurements (all p < 0.05). We found a significant lower signal-to-noise and contrast-to-noise ratio for the HR compared to the LR measurements. The carotid arterial wall dimension measurements of all parameters were significantly lower for the HR compared to the LR measurements. No significant differences were observed between the within-reader and between-reader reproducibility for HR versus LR measurements.
Increasing the in-plane resolution improves the reproducibility of 3T MRI carotid arterial wall dimension measurements. The use of HR imaging will contribute to a reduced sample size needed in intervention trials using MRI scanning of the carotid artery as surrogate marker for atherosclerosis progression.
提高明显的动脉粥样硬化性颈动脉疾病患者的颈动脉3T磁共振成像(MRI)尺寸测量精度。
对31例晚期动脉粥样硬化性颈动脉疾病患者在1个月内进行两次高分辨率(0.25×0.25 mm²;HR)和两次常规使用的低分辨率(0.50×0.50 mm²;LR)颈动脉3T MRI扫描。在一个专门的图像分析程序中,在覆盖动脉粥样硬化斑块的8个切片上手动勾勒颈动脉壁轮廓,确定图像的可重复性以及阅片者内部和阅片者之间的变异性。
我们发现HR测量的总壁体积、平均壁面积和平均壁厚度的组内相关系数显著更高(所有p<0.05)。与LR测量相比,HR的信噪比和对比噪声比显著更低。与LR测量相比,HR的所有参数的颈动脉壁尺寸测量均显著更低。HR与LR测量在阅片者内部和阅片者之间的可重复性方面未观察到显著差异。
提高平面分辨率可提高3T MRI颈动脉壁尺寸测量的可重复性。使用HR成像将有助于减少在以颈动脉MRI扫描作为动脉粥样硬化进展替代标志物的干预试验中所需的样本量。