Evaluation of microbial profile in dental unit waterlines and assessment of antimicrobial efficacy of two treating agents.

作者信息

Mungara J, Dilna N C, Joseph E, Reddy N

机构信息

Ragas Dental College and Hospital, India.

出版信息

J Clin Pediatr Dent. 2013 Summer;37(4):367-71. doi: 10.17796/jcpd.37.4.l6851n9g01223450.

Abstract

OBJECTIVE

The quality of water in a dental unit used for cooling and flushing the high and low speed handpiece, air/water syringes and the scalers is of considerable importance. The present study was carried out to enumerate and identify the microorganisms present in water samples collected from dental unit waterlines of different dental specialty clinics and to find out the efficacy of two treating agents in disinfecting dental unit waterlines.

STUDY DESIGN

Sample included 70 dental unit waterlines from different speciality dental clinics which were checked for microbial contamination. From these dental units 40 units were randomly selected and divided into two groups of 20 each. Group A, treatment was done in 20 dental units with 0.2% Chlorhexidine gluconate solution and Group B, treatment was done in 20 dental units with 10% Povidone iodine solution and the reduction in the microbial levels were assessed. Five dental units were randomly selected and checked the microbial contamination using mineral water sterile distilled water fresh tap water as a water source in the dental unit reservoir bottles. Also from the test group, five from each group were checked for the duration of efficacy of treating agent for one week by analyzing the water samples collected on 3,5 and 7 day intervals.

RESULTS

Most of the identified microorganisms are Gram negative and pseudomonas predominating up to 98.59% of the total isolates. Usage of disinfectants 0.2% Chlorhexidine and 10% Povidone Iodine were found to be very effective in reducing the microbial contamination and 10% Povidone iodine was found to be more efficient (97.13%) and active for a period of 3 days and gradually loosing its efficacy by 7th day. No significant difference were found in microbial contamination of water samples collected from different water outlets such as handpiece outlets, air water syringe outlets, scaler lines.

CONCLUSION

To continue maintaining the sterility of the Dental unit waterlines and to complete the infection control measures adopted in the dental clinics, suitable disinfectants like 0.2% Chlorhexidine on daily basis or 10% Povidone iodine on every 3rd day basis intermittently maintain the sterility of dental unit waterlines it is essential to have a good water source and an effective disinfectant.

摘要

目的

用于冷却和冲洗高速及低速手机、空气/水注射器和洁牙器的牙科设备中的水质至关重要。本研究旨在对从不同牙科专科诊所的牙科设备水线采集的水样中的微生物进行计数和鉴定,并探究两种处理剂对牙科设备水线的消毒效果。

研究设计

样本包括来自不同专科牙科诊所的70条牙科设备水线,对其进行微生物污染检查。从这些牙科设备中随机选取40条,分为两组,每组20条。A组,在20条牙科设备中使用0.2%葡萄糖酸氯己定溶液进行处理;B组,在20条牙科设备中使用10%聚维酮碘溶液进行处理,并评估微生物水平的降低情况。随机选取5条牙科设备,以矿泉水、无菌蒸馏水、新鲜自来水作为牙科设备储水瓶中的水源,检查微生物污染情况。此外,从测试组中,每组随机选取5条,通过分析在第3、5和7天间隔采集的水样,检查处理剂的一周疗效持续时间。

结果

鉴定出的大多数微生物为革兰氏阴性菌,假单胞菌占总分离株的98.59%。发现使用0.2%葡萄糖酸氯己定和10%聚维酮碘消毒剂在减少微生物污染方面非常有效,且10%聚维酮碘更有效(97.13%),并在3天内保持活性,到第7天逐渐失去效力。从不同出水口(如手机出水口、空气/水注射器出水口、洁牙器管线)采集的水样的微生物污染情况未发现显著差异。

结论

为持续保持牙科设备水线的无菌状态并完善牙科诊所采取的感染控制措施,每天使用0.2%葡萄糖酸氯己定或每3天间歇使用10%聚维酮碘等合适的消毒剂来维持牙科设备水线的无菌状态,拥有良好的水源和有效的消毒剂至关重要。