Chaurasia Shyam S, Champakalakshmi Ravi, Angunawela Romesh I, Tan Donald T, Mehta Jodhbir S
Tissue Engineering and Stem Cell Research Group, Singapore Eye Research Institute (SERI), Singapore.
Transl Vis Sci Technol. 2012 Jun 29;1(2):2. doi: 10.1167/tvst.1.2.2. eCollection 2012.
To optimize fibrin glue (FG) spray for ophthalmic surgery using two spray applicators, EasySpray and DuploSpray systems, by varying the distance from point of application and the pressure/flow rate, and to compare the adhesive strength of sutured and sutureless (FG sprayed) conjunctival graft surgery in a rabbit model.
FG was sprayed on a 0.2 mm-thick sheet of paper using EasySpray by variously combining application distances of 2.5, 5, 7.5, and 10 cm with pressures of 10, 15, and 20 psi. DuploSpray was used at the same distances but with varying flow rates of 1 and 2 L/min. Subsequently, FG was sprayed on porcine corneas and FG thickness was analyzed by histology. In addition, adhesive strength of the conjunctival graft (0.5 × 0.5 cm) attached to the rabbit cornea by sutured and sutureless surgery (FG spray) was compared using a tension meter.
Histology measurements revealed that the FG thickness decreased with increases in distance and pressure of spray using the EasySpray applicator on paper and porcine corneal sections. The adhesive strength of the sutured conjunctival graft (41 ± 4.85 [kilopascal] KPa) was found to be higher than the graft attached by spraying (10 ± 2.3 KPa) and the sequential addition of FG (6 ± 0.714 KPa).
The EasySpray applicator formed a uniform spread of FG at a distance-pressure combination of 5 cm and 20 psi. The conjunctival graft attached with sutures had higher adhesive strength compared with grafts glued with a spray applicator. Although the adhesive strength of FG applied through the applicator was similar to the drop-wise sequential technique, the former was more cost effective because more samples could be sprayed compared with the sequential manual technique.
The standardization of the spray system for the application of FG in ophthalmology will provide an economical method for delivering consistent healing results after surgery.
通过改变距应用点的距离以及压力/流速,利用两种喷雾器EasySpray和DuploSpray系统优化用于眼科手术的纤维蛋白胶(FG)喷雾,并在兔模型中比较缝合和无缝合(FG喷雾)结膜移植手术的黏附强度。
使用EasySpray将FG以2.5、5、7.5和10 cm的不同应用距离与10、15和20磅力/平方英寸的压力进行多种组合,喷在一张0.2 mm厚的纸上。以相同距离使用DuploSpray,但流速分别为1和2 L/分钟。随后,将FG喷在猪角膜上,并通过组织学分析FG厚度。此外,使用张力计比较通过缝合和无缝合手术(FG喷雾)附着在兔角膜上的结膜移植片(0.5×0.5 cm)的黏附强度。
组织学测量显示,在纸张和猪角膜切片上使用EasySpray喷雾器时,FG厚度随喷雾距离和压力的增加而减小。发现缝合的结膜移植片的黏附强度(41±4.85 [千帕斯卡] KPa)高于通过喷雾附着的移植片(10±2.3 KPa)和逐滴顺序添加FG的移植片(6±0.714 KPa)。
EasySpray喷雾器在5 cm和20磅力/平方英寸的距离 - 压力组合下形成了均匀分布的FG。与用喷雾器粘贴的移植片相比,用缝线附着的结膜移植片具有更高的黏附强度。尽管通过喷雾器应用FG的黏附强度与逐滴顺序技术相似,但前者更具成本效益,因为与逐滴手动技术相比,可以喷雾更多样本。
眼科中FG应用喷雾系统的标准化将提供一种经济的方法,以在手术后实现一致的愈合效果。