Lauria Andrezza, Rodrigues Danillo Costa, Sato Fábio Ricardo Loureiro, Moreira Roger William Fernandes
Department of Oral and Maxillofacial Surgery, School of Dentistry of Piracicaba, State University of Campinas (UNICAMP), Avenida Limeira, 901, Caixa Postal 52, #191, Areião, 13414-903, Piracicaba, São Paulo, Brazil,
Oral Maxillofac Surg. 2014 Dec;18(4):425-30. doi: 10.1007/s10006-013-0431-4. Epub 2013 Sep 24.
The usage of mini anchors for the treatment of internal derangements of the temporomandibular joint (TMJ) is considered an innovative treatment, and it has presented excellent clinical results in the stabilization of the articular disc. This study aimed to evaluate, through mechanical tensile testing, the resistance of mini anchors for TMJ articular disc repositioning.
Ten thermoplastic polymer poly-ether-ether-ketone (PEEK) mini anchors were tested in artificial polyurethane blocks with a pullout test in a mechanical Instron™ machine, model 4411. The mini anchors were pulled out until failure and the force and displacement were recorded for each specimen. A small standard deviation was noted, which indicated reproducibility.
The findings indicate that Cillen™ mini anchors presented a mean force at failure up to 46.1 N with similar values in the scientific literature. Nevertheless, further studies should be conducted for a comparison of PEEK mini anchors with other commercially available anchors, in addition to studies related to PEEK, a scarcely known material in the dentistry field.
使用微型锚钉治疗颞下颌关节(TMJ)内部紊乱被认为是一种创新疗法,并且在稳定关节盘方面已呈现出优异的临床效果。本研究旨在通过机械拉伸试验评估用于TMJ关节盘重新定位的微型锚钉的抗拉力。
在型号为4411的Instron™ 机械试验机上,通过拔出试验对十个热塑性聚合物聚醚醚酮(PEEK)微型锚钉在人造聚氨酯块中进行测试。将微型锚钉拔出直至失效,并记录每个样本的力和位移。注意到标准偏差较小,这表明具有可重复性。
研究结果表明,Cillen™ 微型锚钉在失效时的平均力高达46.1 N,与科学文献中的值相似。然而,除了对牙科领域鲜为人知的材料PEEK进行相关研究外还应开展进一步研究,以将PEEK微型锚钉与其他市售锚钉进行比较。