Medical University of Vienna, Center for Medical Physics and Biomedical Engineering, Währinger Gürtel 18-20, Vienna 1090, AustriabIntegrated Microsystems Austria GmbH, Viktor Kaplan-Strasse 2/1, Wiener Neustadt 2700, Austria.
J Biomed Opt. 2013 Dec;18(12):121502. doi: 10.1117/1.JBO.18.12.121502.
Low-coherence tissue interferometry is a technique for the depth-resolved measurement of ocular fundus pulsations. Whereas fundus pulsation amplitudes at preselected axial positions can readily be assessed by this method, coupling of the interferometer with a pulse oximeter additionally allows for the reconstruction of the time course of ocular fundus pulsation with respect to the cardiac cycle of the subject. For this purpose, the interferogram resulting from the superposition of waves reflected at the cornea and the ocular fundus is recorded synchronously with the plethysmogram. A new method for evaluating the time course of synthetic interferograms in combination with plethysmograms based on averaging several pulse periods has been developed. This technique allows for the analysis of amplitudes, time courses, and phase differences of fundus pulsations at preselected axial and transversal positions and for creating fundus pulsation movies. Measurements are performed in three healthy emmetropic subjects at angles from 0 deg to 18 deg to the axis of vision. Considerably different time courses, amplitudes, and phases with respect to the cardiac cycle are found at different angles. Data on ocular fundus pulsation obtained with this technique can--among other applications--be used to verify and to improve biomechanical models of the eye.
低相干组织干涉测量是一种用于眼部眼底脉动深度分辨测量的技术。虽然该方法可以很容易地评估在预选轴向位置处的眼底脉动幅度,但通过将干涉仪与脉搏血氧计耦合,还可以重建与主体的心动周期相对应的眼部眼底脉动的时间过程。为此,将在角膜和眼部眼底处反射的波的干涉图与体积描记图同步记录。已经开发出一种新的方法,用于基于平均几个脉搏周期来评估综合干涉图和体积描记图的时间过程。该技术允许在预选的轴向和横向位置分析眼底脉动的幅度、时间过程和相位差,并创建眼底脉动电影。在三个健康的正视眼受试者中,在与视线轴成 0 度至 18 度的角度进行测量。在不同的角度发现了与心动周期相当不同的时间过程、幅度和相位。使用该技术获得的眼部眼底脉动数据可用于验证和改进眼睛的生物力学模型等应用。