Institute of Technical Medicine, Furtwangen University, Jakob-Kienzle Straße 17, D-78054, VS-Schwenningen, Germany. Department of Radiology, University of Munich, Ziemssenstrasse 1, D-80336, Munich, Germany.
Physiol Meas. 2013 Nov;34(11):N107-14. doi: 10.1088/0967-3334/34/11/N107. Epub 2013 Oct 22.
Electrical impedance tomography (EIT) is able to deliver regional information to assess the airway obstruction in patients with cystic fibrosis (CF). In the present study, regional obstruction in CF patients measured by EIT was compared with high resolution computed tomography (HRCT). Five CF patients were routinely scheduled for HRCT examination. EIT measurements were performed on these patients ±2 months during a standard pulmonary function test. The weighted Brody score derived from HRCT, which considers bronchiectasis, mucus plugging, peribronchial thickening, parenchymal opacity and hyperinflation, was calculated from the CT scans acquired at the location of EIT electrodes ±5 cm. Ratios of maximum expiratory flows at 25% and 75% of vital capacity (MEF25/MEF75) with respect to relative impedance change were calculated for regional areas in EIT images. Regional airway obstruction identified in the MEF25/MEF75 maps was similar to that found in CT. Median values of MEF25/MEF75 and weighted Brody score were highly correlated (r(2) = 0.83, P < 0.05). We found that regional obstruction measured by EIT is reliable and may be used as an additional clinical examination tool for CF patients.
电阻抗断层成像(EIT)能够提供区域信息,以评估囊性纤维化(CF)患者的气道阻塞情况。在本研究中,EIT 测量的 CF 患者的区域阻塞情况与高分辨率计算机断层扫描(HRCT)进行了比较。对 5 名 CF 患者进行了 HRCT 检查,在标准肺功能测试期间 ±2 个月对这些患者进行了 EIT 测量。从 HRCT 获得的 CT 扫描中计算了加权 Brody 评分,该评分考虑了支气管扩张、黏液阻塞、支气管周围增厚、肺实质混浊和过度充气,位于 EIT 电极的 ±5cm 处。计算了 EIT 图像中区域的最大呼气流量在 25%和 75%肺活量时与相对阻抗变化的比值(MEF25/MEF75)。在 MEF25/MEF75 图谱中识别的区域性气道阻塞与 CT 中发现的相似。MEF25/MEF75 和加权 Brody 评分的中位数高度相关(r²=0.83,P<0.05)。我们发现,EIT 测量的区域性阻塞是可靠的,可以作为 CF 患者的附加临床检查工具。