The aim of the study was to explore the influence of the measurement plane on regional lung function assessed via electrical impedance tomography (EIT).
The forced vital capacity (FVC) maneuver was prospectively performed in 30 healthy male volunteers. Simultaneously, EIT measurements were conducted at the 3rd, 4th, and 5th intercostal spaces (ICS). The EIT-based spirometry parameters are calculated in a similar manner to their original definitions. The spatial and temporal distributions of the corresponding functional images were assessed and compared among the measurement planes.
All subjects but one were able to perform the FVC maneuver according to the guidelines. Significant differences were found in 67% (6 out of 9) of the EIT-based parameters assessing the spatial and temporal distribution. The fEIT images were most homogeneous at ICS 4 compared to the other two measurement planes, except for the time required for 75% of FVC. The fEIT image FVC distributed toward dorsal regions when the measurement planes moved from ICS 3 to ICS 5, whereas the identified lung areas became smaller.
The spatial and temporal distribution of the regional lung function measured via EIT was influenced by the measurement planes. We recommend adhering to the same measurement plane for before-after comparison. ICS 4 was recommended for the sitting subjects performing lung function testing.
本研究旨在探讨测量平面通过电阻抗断层成像(EIT)对评估局部肺功能的影响。
对30名健康男性志愿者进行前瞻性用力肺活量(FVC)动作测试。同时,在第3、4和5肋间间隙(ICS)进行EIT测量。基于EIT的肺量计参数以与其原始定义类似的方式计算。评估并比较测量平面之间相应功能图像的空间和时间分布。
除一名受试者外,所有受试者均能按照指南完成FVC动作。在评估空间和时间分布的基于EIT的参数中,67%(9项中的6项)存在显著差异。与其他两个测量平面相比,fEIT图像在ICS 4处最为均匀,但达到FVC的75%所需时间除外。当测量平面从ICS 3移至ICS 5时,fEIT图像FVC向背侧区域分布,而识别出的肺区域变小。
通过EIT测量的局部肺功能的空间和时间分布受测量平面影响。我们建议在前后比较时采用相同的测量平面。对于进行肺功能测试的坐位受试者,建议使用ICS 4。