Department of Radiation Oncology, Medical University of Vienna / AKH Wien, Vienna, Austria; Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Vienna, Austria.
Radiother Oncol. 2013 Dec;109(3):437-41. doi: 10.1016/j.radonc.2013.09.020. Epub 2013 Oct 30.
To investigate the differences in treatment plan quality of IMRT and VMAT with and without flattening filter using Pareto optimal fronts, for two treatment sites of different anatomic complexity.
Pareto optimal fronts (POFs) were generated for six prostate and head-and-neck cancer patients by stepwise reduction of the constraint (during the optimization process) of the primary organ-at-risk (OAR). 9-static field IMRT and 360°-single-arc VMAT plans with flattening filter (FF) and without flattening filter (FFF) were compared. The volume receiving 5 Gy or more (V5 Gy) was used to estimate the low dose exposure. Furthermore, the number of monitor units (MUs) and measurements of the delivery time (T) were used to assess the efficiency of the treatment plans.
A significant increase in MUs was found when using FFF-beams while the treatment plan quality was at least equivalent to the FF-beams. T was decreased by 18% for prostate for IMRT with FFF-beams and by 4% for head-and-neck cases, but increased by 22% and 16% for VMAT. A reduction of up to 5% of V5 Gy was found for IMRT prostate cases with FFF-beams.
The evaluation of the POFs showed an at least comparable treatment plan quality of FFF-beams compared to FF-beams for both treatment sites and modalities. For smaller targets the advantageous characteristics of FFF-beams could be better exploited.
利用 Pareto 最优前沿(POF),研究不同解剖结构复杂程度的两个治疗部位在使用和不使用均整滤过器的情况下,调强放疗(IMRT)和容积旋转调强放疗(VMAT)的治疗计划质量差异。
通过逐步减少主要危及器官(OAR)的约束(在优化过程中),为 6 名前列腺癌和头颈部癌症患者生成 Pareto 最优前沿(POF)。比较了 9 野静态调强放疗(IMRT)和 360°单弧容积旋转调强放疗(VMAT),分别使用和不使用均整滤过器(FF 和 FFF)。使用 5Gy 或以上体积(V5Gy)来估计低剂量暴露。此外,还使用了机器跳数(MU)和治疗时间(T)的测量值来评估治疗计划的效率。
使用 FFF 射线时 MU 显著增加,而治疗计划质量至少与 FF 射线相当。对于前列腺 IMRT,FF 射线的 MU 减少了 18%,头颈部病例减少了 4%,而 VMAT 的 MU 增加了 22%和 16%。对于前列腺 IMRT 病例,使用 FFF 射线可将 V5Gy 降低 5%。
通过评估 POF 发现,对于两个治疗部位和治疗方式,FFF 射线的治疗计划质量至少与 FF 射线相当。对于较小的靶区,FFF 射线的优势特性可以更好地发挥。