Kawakami Hiromasa, Miyashita Tetsuya, Yanaizumi Ryota, Mihara Takahiro, Sato Hitoshi, Kariya Takayuki, Mizuno Yusuke, Goto Takahisa
Department of Anesthesiology, Yokohama City University Hospital, Yokohama, Japan.
Technol Health Care. 2013;21(6):581-6. doi: 10.3233/THC-130754.
An unintended bolus is delivered by the syringe pump if intravenous line occlusion is released in an inappropriate manner.
The aim of this study was to measure the amount of flushed fluid when an occlusion is inappropriately released and to assess the effect of different syringe pump settings (flow rate, alarm setting, size of syringe and syringe pump model) on the flushed amount.
After the stopcock was closed, infusions were started with different model syringe pumps (Terufusion® TE312 and TE332S), different syringe sizes or at different alarm settings. After the occlusion alarm sounded, the occlusion was released and the amount of fluid emerging from the stopcock was measured.
The bolus was significantly lower when the alarm was set at a low-pressure setting. The bolus was significantly lower with a 10-ml than a 50-ml syringe. A significant difference was seen only when a 50-ml syringe was used (TE312: 1.99 ± 0.16 ml vs. TE332S: 0.674 ± 0.116 ml, alarm High, p < 0.001).
To minimize the amount of accidentally injected medication, a smaller syringe size and a low alarm setting are important. Using a syringe pump capable of reducing the inadvertently administered bolus may be helpful.
如果以不当方式解除静脉管路阻塞,注射泵会意外推注药物。
本研究旨在测量不当解除阻塞时冲洗出的液体量,并评估不同注射泵设置(流速、警报设置、注射器规格和注射泵型号)对冲洗量的影响。
关闭旋塞后,使用不同型号的注射泵(Terufusion® TE312和TE332S)、不同规格的注射器或在不同警报设置下开始输液。阻塞警报响起后,解除阻塞并测量从旋塞流出的液体量。
将警报设置为低压时,推注量显著降低。使用10 ml注射器时的推注量显著低于50 ml注射器。仅在使用50 ml注射器时观察到显著差异(TE312:1.99 ± 0.16 ml vs. TE332S:0.674 ± 0.116 ml,高警报,p < 0.001)。
为尽量减少意外注射的药物量,较小规格的注射器和低警报设置很重要。使用能够减少无意推注量的注射泵可能会有帮助。