The study was to explore the feasibility of improving cone-beam CT (CBCT) number (corresponding to the Hounsfield units in computed tomography) consistency using a scatter-correction algorithm, with the aim of using CBCT images for treatment planning with density correction. A scatter correction algorithm was applied to a Varian OBI CBCT and an Elekta XVI CBCT, and was evaluated for improving CBCT number consistency. CBCT numbers of phantom materials were compared between images with and without bolus, which introduced additional scatter, and with and without scatter correction processing. It was observed that CBCT numbers were different in the images with and without bolus in both CBCT studies, and the differences were reduced remarkably after scatter-correction processing. Results showed that CBCT number consistency was significantly improved by use of the scatter-correction algorithm.
本研究旨在探索使用散射校正算法提高锥形束 CT(CBCT)号(对应于计算机断层扫描中的亨氏单位)一致性的可行性,以期使用 CBCT 图像进行密度校正的治疗计划。散射校正算法应用于瓦里安 OBI CBCT 和 Elekta XVI CBCT,并对其改善 CBCT 号一致性的能力进行了评估。比较了有和没有模体材料的 bolus(引入额外散射)以及有和没有散射校正处理的图像之间的 CBCT 号。结果表明,在这两项 CBCT 研究中,有和没有 bolus 的图像的 CBCT 号是不同的,并且在散射校正处理后,差异显著减小。结果表明,使用散射校正算法可以显著提高 CBCT 号的一致性。