Phys Med Biol. 2013 Dec 21;58(24):8677-87. doi: 10.1088/0031-9155/58/24/8677. Epub 2013 Nov 22.
Our aim was to measure the performance of desktop magnetic resonance imaging (MRI) systems using specially designed phantoms, by testing imaging parameters and analysing the imaging quality. We designed multifunction phantoms with diameters of 18 and 60 mm for desktop MRI scanners in accordance with the American Association of Physicists in Medicine (AAPM) report no. 28. We scanned the phantoms with three permanent magnet 0.5 T desktop MRI systems, measured the MRI image parameters, and analysed imaging quality by comparing the data with the AAPM criteria and Chinese national standards. Image parameters included: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, signal-to-noise ratio (SNR), and image uniformity. The image parameters of three desktop MRI machines could be measured using our specially designed phantoms, and most parameters were in line with MRI quality control criterion, including: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, image uniformity and slice position accuracy. However, SNR was significantly lower than in some references. The imaging test and quality control are necessary for desktop MRI systems, and should be performed with the applicable phantom and corresponding standards.
我们的目的是使用专门设计的体模来测量台式磁共振成像(MRI)系统的性能,通过测试成像参数和分析成像质量来实现。我们根据美国医学物理学家协会(AAPM)报告 28 号设计了直径为 18 毫米和 60 毫米的多功能体模,用于台式 MRI 扫描仪。我们使用三个永久磁铁 0.5T 台式 MRI 系统对体模进行扫描,测量 MRI 图像参数,并通过将数据与 AAPM 标准和中国国家标准进行比较来分析成像质量。图像参数包括:共振频率、高对比度空间分辨率、低对比度物体探测能力、层厚、几何变形、信噪比(SNR)和图像均匀性。我们专门设计的体模可用于测量三种台式 MRI 机的图像参数,其中大部分参数符合 MRI 质量控制标准,包括:共振频率、高对比度空间分辨率、低对比度物体探测能力、层厚、几何变形、图像均匀性和层位置准确性。然而,SNR 明显低于一些参考文献中的值。台式 MRI 系统需要进行成像测试和质量控制,并且应该使用适用的体模和相应的标准进行。