Suppr超能文献

使用美国放射学会(ACR)磁共振成像(MRI)体模对临床MRI扫描仪进行质量保证:初步结果。

Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results.

作者信息

Chen Chien-Chuan, Wan Yung-Liang, Wai Yau-Yau, Liu Ho-Ling

机构信息

MRI Center, Department of Diagnostic Radiology, Chang Gung Memorial Hospital, Taoyuan, Taiwan.

出版信息

J Digit Imaging. 2004 Dec;17(4):279-84. doi: 10.1007/s10278-004-1023-5.

Abstract

Clinical magnetic resonance imaging (MRI) scanners play an important role in the diagnosis of diseases and management of patient treatment. Quality assurance (QA) of the clinical MRI scanners is mandatory to obtain optimal images in a modern hospital. In this report, the phantom test for the American College of Radiology (ACR) MRI accreditation is used as the essential part of the MRI QA protocols. Seven important assessments of MR image quality are included as follows: geometric accuracy, high-contrast resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low-contrast object detectability. In addition, signal-to-noise ratio and central frequency are monitored as well. The MRI QA procedures were applied to four clinical MRI scanners in our institute twice within 3 months. According to the QA results, the service engineers were more efficient in solving scanners problems when the ACR phantom test was run.

摘要

临床磁共振成像(MRI)扫描仪在疾病诊断和患者治疗管理中发挥着重要作用。在现代医院中,对临床MRI扫描仪进行质量保证(QA)是获得最佳图像的必要条件。在本报告中,美国放射学会(ACR)MRI认证的模体测试被用作MRI QA协议的重要组成部分。包括对MR图像质量的七项重要评估如下:几何精度、高对比度分辨率、层厚精度、层位置精度、图像强度均匀性、信号鬼影百分比和低对比度物体可检测性。此外,还监测信噪比和中心频率。MRI QA程序在3个月内对我院的四台临床MRI扫描仪进行了两次应用。根据QA结果,运行ACR模体测试时,服务工程师在解决扫描仪问题方面效率更高。

相似文献

1
Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results.
J Digit Imaging. 2004 Dec;17(4):279-84. doi: 10.1007/s10278-004-1023-5.
2
An open source automatic quality assurance (OSAQA) tool for the ACR MRI phantom.
Australas Phys Eng Sci Med. 2015 Mar;38(1):39-46. doi: 10.1007/s13246-014-0311-8. Epub 2014 Nov 21.
4
Testing the quality of images for permanent magnet desktop MRI systems using specially designed phantoms.
Phys Med Biol. 2013 Dec 21;58(24):8677-87. doi: 10.1088/0031-9155/58/24/8677. Epub 2013 Nov 22.
5
MRI quality assurance using the ACR phantom in a multi-unit imaging center.
Acta Oncol. 2011 Aug;50(6):966-72. doi: 10.3109/0284186X.2011.582515.
6
Quality assurance for MRI: practical experience.
Br J Radiol. 2000 Apr;73(868):376-83. doi: 10.1259/bjr.73.868.10844863.
10
Automatic Analysis of ACR Phantom Images in MRI.
Curr Med Imaging. 2020;16(7):892-901. doi: 10.2174/1573405615666190903145343.

引用本文的文献

4
Open-source, customizable phantom for low-field magnetic resonance imaging.
MAGMA. 2025 Jun 25. doi: 10.1007/s10334-025-01270-2.
5
Metrology for MRI: the field you've never heard of.
MAGMA. 2025 Mar 19. doi: 10.1007/s10334-025-01238-2.
6
A thermally polarized, dissolved-phase Xe phantom for quality-control and multisite comparisons of gas-exchange imaging.
J Magn Reson. 2025 Feb;371:107829. doi: 10.1016/j.jmr.2025.107829. Epub 2025 Jan 13.
8
Evaluating stereotactic accuracy with patient-specific MRI distortion corrections for frame-based radiosurgery.
J Appl Clin Med Phys. 2024 Sep;25(9):e14472. doi: 10.1002/acm2.14472. Epub 2024 Jul 23.
9
Evaluating Machine Learning-Based MRI Reconstruction Using Digital Image Quality Phantoms.
Bioengineering (Basel). 2024 Jun 15;11(6):614. doi: 10.3390/bioengineering11060614.
10
Phantoms for Quantitative Body MRI: a review and discussion of the phantom value.
MAGMA. 2024 Aug;37(4):535-549. doi: 10.1007/s10334-024-01181-8. Epub 2024 Jun 19.

本文引用的文献

1
Quality assurance for MRI: practical experience.
Br J Radiol. 2000 Apr;73(868):376-83. doi: 10.1259/bjr.73.868.10844863.
3
Automatic quality assessment protocol for MRI equipment.
Med Phys. 1999 Dec;26(12):2693-700. doi: 10.1118/1.598809.
5
The use of power images to perform quantitative analysis on low SNR MR images.
Magn Reson Imaging. 1993;11(7):1051-6. doi: 10.1016/0730-725x(93)90225-3.
6
MRI scanner variability studies using a semi-automated analysis system.
Magn Reson Imaging. 1994;12(7):1089-97. doi: 10.1016/0730-725x(94)91241-n.
7
[Quality control in magnetic resonance for clinical use].
Ann Ist Super Sanita. 1994;30(1):123-37.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验