Chen Chien-Chuan, Wan Yung-Liang, Wai Yau-Yau, Liu Ho-Ling
MRI Center, Department of Diagnostic Radiology, Chang Gung Memorial Hospital, Taoyuan, Taiwan.
J Digit Imaging. 2004 Dec;17(4):279-84. doi: 10.1007/s10278-004-1023-5.
Clinical magnetic resonance imaging (MRI) scanners play an important role in the diagnosis of diseases and management of patient treatment. Quality assurance (QA) of the clinical MRI scanners is mandatory to obtain optimal images in a modern hospital. In this report, the phantom test for the American College of Radiology (ACR) MRI accreditation is used as the essential part of the MRI QA protocols. Seven important assessments of MR image quality are included as follows: geometric accuracy, high-contrast resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low-contrast object detectability. In addition, signal-to-noise ratio and central frequency are monitored as well. The MRI QA procedures were applied to four clinical MRI scanners in our institute twice within 3 months. According to the QA results, the service engineers were more efficient in solving scanners problems when the ACR phantom test was run.
临床磁共振成像(MRI)扫描仪在疾病诊断和患者治疗管理中发挥着重要作用。在现代医院中,对临床MRI扫描仪进行质量保证(QA)是获得最佳图像的必要条件。在本报告中,美国放射学会(ACR)MRI认证的模体测试被用作MRI QA协议的重要组成部分。包括对MR图像质量的七项重要评估如下:几何精度、高对比度分辨率、层厚精度、层位置精度、图像强度均匀性、信号鬼影百分比和低对比度物体可检测性。此外,还监测信噪比和中心频率。MRI QA程序在3个月内对我院的四台临床MRI扫描仪进行了两次应用。根据QA结果,运行ACR模体测试时,服务工程师在解决扫描仪问题方面效率更高。