Diagnostic performance of low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT using the 530c CZT camera: quantitative vs visual analysis.

作者信息

Esteves Fabio P, Galt James R, Folks Russell D, Verdes Liudmila, Garcia Ernest V

机构信息

Department of Radiology, Emory University School of Medicine, 1364 Clifton Road, NE, Atlanta, GA, 30322, USA,

出版信息

J Nucl Cardiol. 2014 Feb;21(1):158-65. doi: 10.1007/s12350-013-9827-7. Epub 2013 Nov 28.

Abstract

BACKGROUND

We set out to develop normal databases and prospectively validate abnormality criteria for a low-dose Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera.

METHODS

All patients received 6 mCi rest/20 mCi stress doses of Tc-99m tetrofosmin. Rest and stress images were obtained over 7-9 and 5-7 minutes according to the chest size. Low-dose CT of the chest was obtained on a standalone CT scanner. Forty patients with very low likelihood (LLK) of coronary artery disease (CAD) were used to define the normal count distributions. The abnormality criteria were prospectively validated in 55 patients who had coronary angiography and in 40 patients with LLK of CAD.

RESULTS

The results for quantitative non-attenuation-corrected (AC) and AC analysis and visual analysis were as follows: sensitivity of 79%, 85%, and 92% (P = NS) and specificity of 44%, 75%, and 56% (P = NS), respectively. The normalcy rates for quantitative non-AC and AC analyses and visual analysis were 95%, 98%, and 98% (P = NS).

CONCLUSIONS

We have developed non-AC and AC normal databases for low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera. The per-patient diagnostic performance of quantitative analyses is not significantly different from visual analysis by an experienced reader.

摘要

背景

我们着手开发正常数据库,并前瞻性地验证使用530c CZT相机的低剂量锝-99m替曲膦心肌灌注单光子发射计算机断层扫描(SPECT)协议的异常标准。

方法

所有患者接受6毫居里静息/20毫居里负荷剂量的锝-99m替曲膦。根据胸部大小,在7 - 9分钟和5 - 7分钟内分别获取静息和负荷图像。在一台独立的CT扫描仪上进行胸部低剂量CT扫描。40例冠状动脉疾病(CAD)可能性极低(LLK)的患者用于定义正常计数分布。在55例接受冠状动脉造影的患者和40例CAD可能性为LLK的患者中前瞻性地验证异常标准。

结果

定量非衰减校正(AC)和AC分析以及视觉分析的结果如下:敏感性分别为79%、85%和92%(P = 无显著性差异),特异性分别为44%、75%和56%(P = 无显著性差异)。定量非AC和AC分析以及视觉分析的正常率分别为95%、98%和98%(P = 无显著性差异)。

结论

我们使用530c CZT相机为低剂量静息/负荷锝-99m替曲膦心肌灌注SPECT协议开发了非AC和AC正常数据库。定量分析的每位患者诊断性能与经验丰富的阅片者的视觉分析无显著差异。