Dias M S, Brancaccio F, Toledo F, Koskinas M F
Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP, Av. Prof. Lineu Prestes 2242, 05508-000 São Paulo, SP, Brazil.
Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP, Av. Prof. Lineu Prestes 2242, 05508-000 São Paulo, SP, Brazil.
Appl Radiat Isot. 2014 May;87:126-31. doi: 10.1016/j.apradiso.2013.11.001. Epub 2013 Nov 13.
The procedure for determining the (67)Ga disintegration rate by a primary method is described. The proposed triple 4πβ-γ coincidence system consists of a thin window gas-flow 4π proportional counter (PC) coupled to a NaI(Tl) scintillator and a HPGe crystal. Independent pulse height and occurrence time information is provided for the three detector outputs by means of a Software Coincidence System. Separate spectrometry measurements with a n-type reversed electrode coaxial Ge detector (REGe) were performed for obtaining gamma-ray emission probabilities per decay. Accurate values of disintegration rate, gamma-ray emission probabilities and the metastable half-life were achieved.