* Audiology Department, Hearing Centre, Croydon University Hospital , Croydon , UK.
Int J Audiol. 2014 Feb;53(2):132-7. doi: 10.3109/14992027.2013.843027. Epub 2013 Dec 10.
To estimate bone-conduction stimulus level corrections by testing the auditory brainstem response (ABR) of normally-hearing newborns. The stimuli used were low frequency tone pips calibrated to reference levels derived from ISO 389 values.
Tone pips were presented via supra-aural earphones and a B71 Radioear bone vibrator at 0.5 or 1 kHz. ABR thresholds from both transducers were compared at each frequency.
twenty-seven newborn hearing screening referrals (33 ears) who passed an ABR discharge criterion at 4 kHz.
Median air- and bone-conduction ABR threshold differences were 30 dB at 0.5 kHz and 20 dB at 1 kHz.
The 0.5 kHz data from this study and previous studies were compared. Previous studies suggested lower figures for the bone-conduction stimulus level correction. Likely sources of this discrepancy are discussed. The average 0.5 kHz bone-conduction correction value for infants < 3 months old is about 28 dB. The correction for 1 kHz is 20 dB. We recommend that calibration reference levels used in this study be adopted and that appropriate corrections be applied to bone conduction ABR thresholds in infants < 3 months old before calculation of any air-bone gap and subsequent clinical interpretation.
通过测试正常听力新生儿的听脑干反应(ABR)来估计骨导刺激水平的校正。所使用的刺激为低频音调脉冲,其校准参考水平来自 ISO 389 值。
通过耳罩式耳机和 B71 Radioear 骨振器以 0.5 或 1 kHz 的频率呈现音调脉冲。在每个频率下,比较两个换能器的 ABR 阈值。
通过 4 kHz ABR 放电标准的 27 名新生儿听力筛查转诊(33 只耳朵)。
在 0.5 kHz 时,气导和骨导 ABR 阈值差异的中位数为 30 dB,在 1 kHz 时为 20 dB。
比较了这项研究和以前研究中的 0.5 kHz 数据。以前的研究表明,骨导刺激水平校正的数值较低。讨论了这种差异的可能来源。<3 个月大婴儿的平均 0.5 kHz 骨导校正值约为 28 dB。1 kHz 的校正值为 20 dB。我们建议采用本研究中使用的校准参考水平,并在计算任何气骨间隙和随后的临床解释之前,对<3 个月大的婴儿的骨导 ABR 阈值进行适当的校正。