Department of Biomedical Engineering, Marquette University, Milwaukee, Wisconsin 53233.
Med Phys. 2013 Dec;40(12):121905. doi: 10.1118/1.4829521.
This study investigated the effects of tilted-gantry acquisition on image noise and glandular breast dose in females during cardiac computed tomography (CT) scans. Reducing the dose to glandular breast tissue is important due to its high radiosensitivity and limited diagnostic significance in cardiac CT scans.
Tilted-gantry acquisition was investigated through computer simulations and experimental measurements. Upon IRB approval, eight voxelized phantoms were constructed from previously acquired cardiac CT datasets. Monte Carlo simulations quantified the dose deposited in glandular breast tissue over a range of tilt angles. The effects of tilted-gantry acquisition on breast dose were measured on a clinical CT scanner (CT750HD, GE Healthcare) using an anthropomorphic phantom with MOSFET dosimeters in the breast regions. In both simulations and experiments, scans were performed at gantry tilt angles of 0°-30°, in 5° increments. The percent change in breast dose was calculated relative to the nontilted scan for all tilt angles. The percent change in noise standard deviation due to gantry tilt was calculated in all reconstructed simulated and experimental images.
Tilting the gantry reduced the breast dose in all simulated and experimental phantoms, with generally greater dose reduction at increased gantry tilts. For example, at 30° gantry tilt, the dosimeters located in the superior, middle, and inferior breast regions measured dose reductions of 74%, 61%, and 9%, respectively. The simulations estimated 0%-30% total breast dose reduction across the eight phantoms and range of tilt angles. However, tilted-gantry acquisition also increased the noise standard deviation in the simulated phantoms by 2%-50% due to increased pathlength through the iodine-filled heart. The experimental phantom, which did not contain iodine in the blood, demonstrated decreased breast dose and decreased noise at all gantry tilt angles.
Tilting the gantry reduced the dose to the breast, while also increasing noise standard deviation. Overall, the noise increase outweighed the dose reduction for the eight voxelized phantoms, suggesting that tilted gantry acquisition may not be beneficial for reducing breast dose while maintaining image quality.
本研究旨在探讨心脏 CT 扫描中倾斜机架采集对女性图像噪声和乳腺剂量的影响。降低乳腺组织的剂量很重要,因为它在心脏 CT 扫描中对乳腺的辐射敏感性高,且诊断意义有限。
通过计算机模拟和实验测量研究了倾斜机架采集。在获得机构审查委员会的批准后,从之前获得的心脏 CT 数据集构建了 8 个体素化体模。蒙特卡罗模拟量化了在一系列倾斜角度下沉积在乳腺组织中的剂量。在临床 CT 扫描仪(GE Healthcare 的 CT750HD)上使用带有乳腺区域 MOSFET 剂量计的人体模型,测量倾斜机架采集对乳腺剂量的影响。在模拟和实验中,在 0°-30°的机架倾斜角度范围内,以 5°的增量进行扫描。对于所有倾斜角度,计算相对于非倾斜扫描的乳腺剂量的百分比变化。在所有重建的模拟和实验图像中,计算由于机架倾斜引起的噪声标准偏差的百分比变化。
倾斜机架降低了所有模拟和实验体模中的乳腺剂量,随着机架倾斜度的增加,剂量降低幅度通常更大。例如,在 30°机架倾斜时,位于乳腺上、中、下区域的剂量计测量到的剂量减少分别为 74%、61%和 9%。模拟估计了八个体模和倾斜角度范围内 0%-30%的总乳腺剂量减少。然而,由于碘填充心脏的路径长度增加,倾斜机架采集也使模拟体模中的噪声标准偏差增加了 2%-50%。不含血液碘的实验体模在所有机架倾斜角度下均显示出乳腺剂量降低和噪声降低。
倾斜机架降低了乳腺剂量,同时也增加了噪声标准偏差。总体而言,对于八个体模,噪声增加超过了剂量减少,这表明倾斜机架采集可能不利于在保持图像质量的同时降低乳腺剂量。