Ultrasound-guided needle handling using a guidance positioning system in a phantom.

机构信息

Radboud University Nijmegen Medical Centre, Nijmegen, the Netherlands.

出版信息

Anaesthesia. 2014 Jan;69(1):24-31. doi: 10.1111/anae.12461. Epub 2013 Nov 9.

Abstract

The SonixGPS™ needle guidance positioning system provides navigation assistance to facilitate needle handling during ultrasound-guided procedures. Each of 20 inexperienced nurse anaesthetists performed 12 different ultrasound-guided tasks in a porcine phantom. Using both in-plane and out-of-plane approaches, they inserted a needle and made contact with metal rods at depths of 2, 4 and 6 cm. We compared their performances without and with navigation as paired observations. Using the out-of-plane approach, navigation yielded shorter execution times (26 s vs 14 s, respectively; p = 0.01) and fewer needle repositionings (8 vs 3, respectively; p = 0.001). Using the in-plane approach, the needle was more visible with navigation assistance: 24% vs 52% of execution time, respectively (95% CI: 44%-12%; p = 0.0025). Better needle visibility was associated with shorter execution times and fewer needle repositionings. Combining ultrasound-guided techniques with the needle guidance positioning system may reduce tissue manipulation, thus improving patient comfort and safety.

摘要

SonixGPS™ 针引导定位系统为超声引导手术中的针操作提供导航辅助。20 名没有经验的护士麻醉师在猪模型上进行了 12 项不同的超声引导任务。他们使用平面内和平面外方法,将针插入并接触到深度为 2、4 和 6 厘米的金属棒。我们将无导航和有导航的情况下的表现作为配对观察进行比较。使用平面外方法,导航使执行时间更短(分别为 26 秒和 14 秒;p = 0.01),重新定位针的次数更少(分别为 8 次和 3 次;p = 0.001)。使用平面内方法,导航辅助下针更可见:分别为执行时间的 24%和 52%(95%CI:44%-12%;p = 0.0025)。更好的针可见度与更短的执行时间和更少的针重新定位有关。将超声引导技术与针引导定位系统相结合,可能会减少组织操作,从而提高患者的舒适度和安全性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验