评估三种市售超低熔陶瓷系统与II级钛之间弯曲粘结强度的体外研究。

In-vitro investigation to evaluate the flexural bond strengths of three commercially available ultra low fusing ceramic systems to Grade II Titanium.

作者信息

Mabrurkar Vijay, Habbu Nitin, Hashmi Sayed Wahhiuddin, Musani Smita, Joshi Nikhil

机构信息

Department of Prosthodontics, M A Rangoonwala College of Dental Sciences & Research Centre, Pune, Maharashtra, India.

出版信息

J Int Oral Health. 2013 Oct;5(5):101-7. Epub 2013 Oct 26.

Abstract

BACKGROUND

This in-vitro investigation was designed to compare the flexural bond strengths of three commercially available ultra low fusing ceramic systems to Grade II Titanium and evaluate the type of bond failure through Scanning Electron Microscopy (SEM) and Energy Dispersion X ray Spectrum (EDS).

MATERIALS & METHODS: Sixty patterns of auto polymerizing resin of dimensions 25.1 mm X 3.1mm X 0.6mm each were fabricated from a stainless steel die. Titanium casting for all the samples was done in a Titanium casting machine. Ten samples were allotted to each of three groups randomly. Ceramic build up was done step by step using the manufacturers' instructions. Flexural Bond strength tests for the samples were performed by using a 3-point bending test on a Universal testing machine in compliance with Deutsches Institut für Normung (DIN) 13.927 draft. After the tests were completed, three samples, one from each group were selected randomly for the subsequent Scanning Electronic Microscopy (SEM) studies to characterize the type and morphology of the fracture in representative specimens.

RESULTS

The maximum load to fracture the porcelain was recorded for each specimen. All the means of the three groups were compared by one-way Analysis of Variance (ANOVA) and it was found that Group I & Group III had significantly higher bond strength values as compared with the Group II (P< 0.001). All the 10 samples of Group I & III gave bond strength values above the standard values of 25 MPa. There was statistically significant difference in the bond strengths between Group I & Group III (t = 2.76 and P< 0.05), between Group I & II (t = 5.09 and P< 0.001) and between Group II & Group III (t = 13.28 and P< 0.001). SEM studies revealed occurrence of cohesive type fractures in the ceramic body of samples belonging to Groups I & III, while there was adhesive failure at ceramo-metal junction of samples belonging to Group II. EDS Analysis supported the findings of SEM studies.

CONCLUSION

All the three ceramic systems fared better than the minimal recommended values stated by ISO 9693:1999 of 25MPa of which Orotig TiKrom was rated the best with values of 54.69 MPa. Vita TitanKeramik ranked second with values of 45.12MPa and the least values were obtained with Noritake Ti-22 with values of 27.76. The bond failure was predominantly cohesive in nature in case of Vita TitanKeramik and Orotig TiKrom; whereas adhesive failure was noticed in case of Noritake Ti-22. How to cite this article: Mabrurkar V, Habbu N, Hashmi SW, Musani S, Joshi N. In-vitro investigation to evaluate the flexural bond strengths of three commercially available ultra low fusing ceramic systems to Grade II Titanium. J Int Oral Health 2013;5(5):101-7.

摘要

背景

本体外研究旨在比较三种市售超低熔陶瓷系统与二级钛的弯曲粘结强度,并通过扫描电子显微镜(SEM)和能量色散X射线光谱(EDS)评估粘结失效的类型。

材料与方法

用不锈钢模具制作60个尺寸为25.1mm×3.1mm×0.6mm的自凝树脂模型。所有样品的钛铸件均在钛铸造机中完成。将10个样品随机分配到三个组中。按照制造商的说明逐步进行陶瓷堆积。使用万能试验机按照德国标准化学会(DIN)13.927草案进行三点弯曲试验,对样品进行弯曲粘结强度测试。测试完成后,从每组中随机选择三个样品进行后续的扫描电子显微镜(SEM)研究,以表征代表性试样中骨折的类型和形态。

结果

记录每个试样使瓷器断裂的最大载荷。通过单向方差分析(ANOVA)比较三组的所有平均值,发现与第二组相比,第一组和第三组具有明显更高的粘结强度值(P<0.001)。第一组和第三组的所有10个样品的粘结强度值均高于25MPa的标准值。第一组和第三组之间(t = 2.76,P<0.05)、第一组和第二组之间(t = 5.09,P<0.001)以及第二组和第三组之间(t = 13.28,P<0.001)的粘结强度存在统计学上的显著差异。SEM研究显示,属于第一组和第三组的样品的陶瓷体中出现内聚型断裂,而属于第二组的样品在陶瓷-金属结合处出现粘结失效。EDS分析支持了SEM研究的结果。

结论

所有三种陶瓷系统的表现均优于ISO 9693:1999规定的最低推荐值25MPa,其中奥罗蒂格TiKrom的表现最佳,值为54.69MPa。维他泰坦陶瓷排名第二,值为45.12MPa,而则武Ti-22的值最低,为27.76MPa。维他泰坦陶瓷和奥罗蒂格TiKrom的粘结失效在本质上主要是内聚性的;而则武Ti-22则出现粘结失效。如何引用本文:Mabrurkar V, Habbu N, Hashmi SW, Musani S, Joshi N. 体外研究评估三种市售超低熔陶瓷系统与二级钛的弯曲粘结强度。《国际口腔健康杂志》2013;5(5):101-7。

相似文献

本文引用的文献

2
Porosity and accuracy of multiple-unit titanium castings.多单元钛铸件的孔隙率与精度
J Prosthet Dent. 1995 Jun;73(6):534-41. doi: 10.1016/s0022-3913(05)80112-x.
5
New approach for evaluating metal-porcelain interfacial bonding.
Int J Prosthodont. 1999 Nov-Dec;12(6):547-52.
6
Fit of titanium and a base metal alloy metal-ceramic crown.钛与贱金属合金金属烤瓷冠的适合性。
J Prosthet Dent. 2000 Mar;83(3):314-8. doi: 10.1016/s0022-3913(00)70134-x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验