Ashokkumar P, Sahoo B K, Raman Anand, Mayya Y S
Radiation Safety Systems Division, Bhabha Atomic Research Centre, Trombay, Mumbai-400085, India.
J Radiol Prot. 2014 Mar;34(1):149-60. doi: 10.1088/0952-4746/34/1/149. Epub 2013 Dec 13.
This paper discusses the development and characterisation of a portable and highly sensitive continuous radon monitor (CRM) based on an array of in-house developed silicon PIN diode detectors. The development of this system was initiated in view of the limitations of the available similar radon measurement systems with regards to low sensitivity. The system utilises a hemispherical metal chamber (1 L capacity) for active air sampling. A quantitative estimation of radon concentration is carried out through alpha spectroscopy of electro-deposited (222)Rn decay products on the detector surface. The system was successfully tested and characterised in laboratory conditions. The characterisation experiments included optimisation of sensitivity, calibration with respect to linearity and a study of the influence of humidity on its performance. The novel PIN diode array design yields a high sensitivity of 1.76 ± 0.003 counts h(-1)/(Bq m(-3)) at a relative humidity level of 10% in the sampled air, which is more than two times as high as that reported for similar commercial systems. This instrument displayed a minimum detectable activity level of 0.80 Bq m(-3).
本文讨论了一种基于一系列自行研制的硅PIN二极管探测器的便携式高灵敏度连续氡监测仪(CRM)的开发与特性。鉴于现有类似氡测量系统在低灵敏度方面的局限性,启动了该系统的开发。该系统采用一个半球形金属腔(容量为1升)进行主动空气采样。通过对探测器表面电沉积的(222)Rn衰变产物进行α能谱分析,对氡浓度进行定量估计。该系统在实验室条件下成功进行了测试和特性表征。特性表征实验包括灵敏度优化、线性校准以及湿度对其性能影响的研究。新颖的PIN二极管阵列设计在采样空气中相对湿度为10%时,产生了1.76±0.003计数·小时-1/(贝克勒尔·立方米-3)的高灵敏度,这比类似商业系统报道的灵敏度高出两倍多。该仪器的最低可探测活度水平为0.80贝克勒尔·立方米-3。