Singh Kanwarpal, Haydari Nour, Brunette Isabelle, Costantino Santiago
Centre de Recherche de l'Hôpital Maisonneuve-Rosemont, Montréal, Québec, Canada ; Department of Ophthalmology, Faculty of Medicine, Université de Montréal, Montréal, Québec, Canada.
PLoS One. 2013 Dec 5;8(12):e83185. doi: 10.1371/journal.pone.0083185. eCollection 2013.
Corneal grafts for Descemet's Stripping Automated Endothelial Keratoplasty are commonly prepared using mechanical microkeratomes. However, the cuts produced in such way render corneal lenticules that are thinner centrally than peripherally, thus inducing a hyperopic shift. Here we describe a novel device for preparing donor corneal grafts, in which a single low-energy femtosecond laser system is used as both a light source for optical coherence tomography and for cutting the graft illuminating from the endothelial side. The same laser is first utilized to obtain three-dimensional optical coherence tomography images of the donor tissue for guiding the dissection and obtaining grafts of uniform thickness with no applanation or contact. This device allows an optimal procedure for preparing consistently thin posterior grafts for transplantation.
用于Descemet膜剥除自动内皮角膜移植术的角膜移植物通常使用机械微型角膜刀制备。然而,以这种方式产生的切口会使角膜透镜中央比周边更薄,从而导致远视移位。在此,我们描述一种用于制备供体角膜移植物的新型装置,其中单个低能量飞秒激光系统既用作光学相干断层扫描的光源,又用于从内皮侧照射切割移植物。首先使用同一激光获取供体组织的三维光学相干断层扫描图像,以指导解剖并获得厚度均匀、无压平或接触的移植物。该装置允许采用最佳程序来制备始终如一的薄型后段移植物用于移植。