Multicenter harmonization of 89Zr PET/CT performance.

作者信息

Makris Nikolaos E, Boellaard Ronald, Visser Eric P, de Jong Johan R, Vanderlinden Bruno, Wierts Roel, van der Veen Berlinda J, Greuter Henri J N M, Vugts Danielle J, van Dongen Guus A M S, Lammertsma Adriaan A, Huisman Marc C

机构信息

Department of Radiology and Nuclear Medicine, VU University Medical Center, Amsterdam, The Netherlands.

出版信息

J Nucl Med. 2014 Feb;55(2):264-7. doi: 10.2967/jnumed.113.130112. Epub 2013 Dec 19.

Abstract

UNLABELLED

This study investigated the feasibility of quantitative accuracy and harmonized image quality in (89)Zr-PET/CT multicenter studies.

METHODS

Five PET/CT scanners from 3 vendors were included. (89)Zr activity was measured in a central dose calibrator before delivery. Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were determined from a cross calibration experiment. Image quality was assessed from recovery coefficients derived from different volume-of-interest (VOI) methods (VOI A 50%, based on a 3-dimensional isocontour at 50% of the maximum voxel value with local background correction; VOI Max, based on the voxel with the highest uptake; and VOI 3Dpeak, based on a spheric VOI of 1.2-cm diameter positioned so as to maximize the enclosed average). PET images were analyzed before and after postreconstruction smoothing, applied to match image noise.

RESULTS

PET/CT accuracy and image noise ranged from -3% to 10% and from 13% to 22%, respectively. VOI 3Dpeak produced the most reproducible recovery coefficients. After calibration of the local dose calibrator to the central dose calibrator, differences between the local activity assays were within 6%.

CONCLUSION

This study showed that quantitative accuracy and harmonized image quality can be reached in (89)Zr PET/CT multicenter studies.

摘要

未标注

本研究调查了⁸⁹Zr-PET/CT多中心研究中定量准确性和图像质量一致性的可行性。

方法

纳入了来自3家供应商的5台PET/CT扫描仪。在交付前,在中央剂量校准仪中测量⁸⁹Zr活度。局部活度测定基于体积以及局部剂量校准仪。通过交叉校准实验确定准确性和图像噪声。根据不同感兴趣体积(VOI)方法得出的恢复系数评估图像质量(VOI A 50%,基于最大体素值50%处的三维等轮廓并进行局部背景校正;VOI Max,基于摄取量最高的体素;VOI 3Dpeak,基于直径为1.2 cm的球形VOI,其位置可使封闭平均值最大化)。在重建后平滑前后分析PET图像,以匹配图像噪声。

结果

PET/CT的准确性和图像噪声分别在-3%至10%和13%至22%之间。VOI 3Dpeak产生的恢复系数最具可重复性。将局部剂量校准仪校准到中央剂量校准仪后,局部活度测定之间的差异在6%以内。

结论

本研究表明,在⁸⁹Zr PET/CT多中心研究中可以实现定量准确性和图像质量的一致性。