Makris Nikolaos E, Boellaard Ronald, Visser Eric P, de Jong Johan R, Vanderlinden Bruno, Wierts Roel, van der Veen Berlinda J, Greuter Henri J N M, Vugts Danielle J, van Dongen Guus A M S, Lammertsma Adriaan A, Huisman Marc C
Department of Radiology and Nuclear Medicine, VU University Medical Center, Amsterdam, The Netherlands.
J Nucl Med. 2014 Feb;55(2):264-7. doi: 10.2967/jnumed.113.130112. Epub 2013 Dec 19.
This study investigated the feasibility of quantitative accuracy and harmonized image quality in (89)Zr-PET/CT multicenter studies.
Five PET/CT scanners from 3 vendors were included. (89)Zr activity was measured in a central dose calibrator before delivery. Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were determined from a cross calibration experiment. Image quality was assessed from recovery coefficients derived from different volume-of-interest (VOI) methods (VOI A 50%, based on a 3-dimensional isocontour at 50% of the maximum voxel value with local background correction; VOI Max, based on the voxel with the highest uptake; and VOI 3Dpeak, based on a spheric VOI of 1.2-cm diameter positioned so as to maximize the enclosed average). PET images were analyzed before and after postreconstruction smoothing, applied to match image noise.
PET/CT accuracy and image noise ranged from -3% to 10% and from 13% to 22%, respectively. VOI 3Dpeak produced the most reproducible recovery coefficients. After calibration of the local dose calibrator to the central dose calibrator, differences between the local activity assays were within 6%.
This study showed that quantitative accuracy and harmonized image quality can be reached in (89)Zr PET/CT multicenter studies.