Suppr超能文献

Ceramic femoral component fracture in total knee arthroplasty: an analysis using fractography, fourier-transform infrared microscopy, contact radiography and histology.

作者信息

Krueger Alexander P, Singh Gurpal, Beil Frank Timo, Feuerstein Bernd, Ruether Wolfgang, Lohmann Christoph H

机构信息

Department of Orthopaedic Surgery, Otto-von-Guericke University, Magdeburg, Germany.

Department of Orthopaedic Surgery, Otto-von-Guericke University, Magdeburg, Germany; University Orthopaedics, Hand and Reconstructive Microsurgery Cluster, National, University Health System, Singapore, Singapore.

出版信息

J Arthroplasty. 2014 May;29(5):1001-4. doi: 10.1016/j.arth.2013.11.003. Epub 2013 Nov 9.

Abstract

Ceramic components in total knee arthroplasty (TKA) are evolving. We analyze the first case of BIOLOX delta ceramic femoral component fracture. A longitudinal midline fracture in the patellar groove was present, with an intact cement mantle and no bony defects. Fractographic analysis with laser scanning microscopy and white light interferometry showed no evidence of arrest lines, hackles, wake hackles, material flaws, fatigue or crack propagation. Analysis of periprosthetic tissues with Fourier-transform infrared (FT-IR) microscopy, contact radiography, histology, and subsequent digestion and high-speed centrifugation did not show ceramic debris. A macrophage-dominated response was present around polyethylene debris. We conclude that ceramic femoral component failure in this case was related to a traumatic event. Further research is needed to determine the suitability of ceramic components in TKA.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验