Oonishi Hironobu, Ueno Masaru, Kim Sok Chol, Oonishi Hiroyuki, Iwamoto Mikio, Kyomoto Masayuki
H. Oonishi Memorial Joint Replacement Institute, Tominaga Hospital, Naniwa-ku, Osaka, Japan.
J Arthroplasty. 2009 Apr;24(3):374-82. doi: 10.1016/j.arth.2007.10.021. Epub 2008 Jun 4.
The present study aimed to determine the effect of femoral component materials and sterilization methods on wear properties of total knee prostheses by using a knee simulator test and retrieval analysis. The simulator test revealed that ultrahigh molecular weight polyethylene (UHMWPE) inserts had remarkably lower wear against the ceramic femoral component than against the Co-Cr femoral component. However, the retrieval study revealed no significant difference in the linear wear between the former and the latter. The alumina ceramic/UHMWPE insert combination showed a mild wear. However, whether cross-linking by gamma-ray sterilization reduces wear remained unconfirmed. In contrast, oxidative degradation and/or delamination was confirmed. Thus, we conclude that alumina ceramic/ethylene oxide gas-sterilized UHMWPE insert in a total knee prosthesis might exhibit a good wear resistance.