Suppr超能文献

6兆伏与15兆伏光子能量计划用于宫颈癌调强放射治疗(IMRT)的剂量学分析。

A dosimetric analysis of 6 MV versus 15 MV photon energy plans for intensity modulated radiation therapy (IMRT) of carcinoma of cervix.

作者信息

Tyagi Atul, Supe Sanjay S, Singh Man P

机构信息

Department of Radiation Oncology, BLK Memorial Hospital, New Delhi, India ; Department of Physics, MMH College, Ghaziabad, UP, India.

Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bangalore, Karnataka, India.

出版信息

Rep Pract Oncol Radiother. 2010 Sep 22;15(5):125-31. doi: 10.1016/j.rpor.2010.08.002. eCollection 2010.

Abstract

BACKGROUND

Intensity modulated radiotherapy (IMRT) is being used to treat carcinoma of cervix (Ca Cx). Integral dose to normal tissue and increased leakage are the concern about IMRT. 6 MV photon beam is a good choice of energy for Ca Cx IMRT treatment.

AIM

The objective of this study was to compare intensity modulated radiotherapy (IMRT) plans generated by 6 MV and 15 MV photon energies for carcinoma of cervix (Ca Cx) with regards to dosimetric parameters of planning target volume (PTV) and organs at risk (OAR), homogeneity index (HI), conformity index at 98% level (CI 98%), integral dose to normal tissue (NTID) and total number of monitor units (MUs).

MATERIAL AND METHODS

A cohort of 16 patients was selected for this study. All patients were to receive a dose of 50 Gy in 25 fractions. IMRT plans were generated for both energies using same dose-volume constraints.

RESULTS

Our results show a comparable coverage of planning target volume (PTV) for both energies. Volume of PTV receiving a prescription dose is 97.8 ± 0.5% and 98.8 ± 0.4% for the 6 MV and the 15 MV plans. Volume of PTV receiving a dose of 107% is 4.4 ± 7.8% and 16.1 ± 22.2%. Bladder and rectum mean doses for the 6 MV and the 15 MV photon plans were 39.8 ± 3.0 Gy and 40.0 ± 3.2 Gy, and 35.8 ± 3.1 Gy and 36.0 ± 3.1 Gy, respectively. Homogeneity index (HI) for both energies was 1.04. The conformity indices at 98% isodose (CI 98%) were 1.3 ± 0.1 and 1.4 ± 0.1 for 6 MV and 15 MV photon plans, respectively.

CONCLUSIONS

We conclude that a 6 MV photon is a good choice for Ca Cx IMRT as it produces a highly conformal, homogeneous plan with superior target coverage and better OAR sparing.

摘要

背景

调强放射治疗(IMRT)正用于治疗宫颈癌(Ca Cx)。正常组织的积分剂量和泄漏增加是调强放射治疗所关注的问题。6兆伏光子束是宫颈癌调强放射治疗的理想能量选择。

目的

本研究的目的是比较6兆伏和15兆伏光子能量产生的调强放射治疗(IMRT)计划在宫颈癌(Ca Cx)的计划靶区(PTV)和危及器官(OAR)的剂量学参数、均匀性指数(HI)、98%剂量水平的适形指数(CI 98%)、正常组织的积分剂量(NTID)和监测单位总数(MUs)方面的差异。

材料与方法

本研究选取了16例患者。所有患者均接受25次分割、总剂量50 Gy的治疗。使用相同的剂量体积限制为两种能量生成调强放射治疗计划。

结果

我们的结果显示两种能量对计划靶区(PTV)的覆盖相当。6兆伏和15兆伏计划中接受处方剂量的PTV体积分别为97.8±0.5%和98.8±0.4%。接受107%剂量的PTV体积分别为4.4±7.8%和16.1±22.2%。6兆伏和15兆伏光子计划中膀胱和直肠的平均剂量分别为39.8±3.0 Gy和40.0±3.2 Gy,以及35.8±3.1 Gy和36.0±3.1 Gy。两种能量的均匀性指数(HI)均为1.04。6兆伏和15兆伏光子计划在98%等剂量线的适形指数(CI 98%)分别为1.3±0.1和1.4±0.1。

结论

我们得出结论,6兆伏光子是宫颈癌调强放射治疗的理想选择,因为它能产生高度适形、均匀的计划,具有更好的靶区覆盖和更好的危及器官保护。

相似文献

1
A dosimetric analysis of 6 MV versus 15 MV photon energy plans for intensity modulated radiation therapy (IMRT) of carcinoma of cervix.
Rep Pract Oncol Radiother. 2010 Sep 22;15(5):125-31. doi: 10.1016/j.rpor.2010.08.002. eCollection 2010.
2
Dosimetric influence of photon beam energy and number of arcs on volumetric modulated arc therapy in carcinoma cervix: A planning study.
Rep Pract Oncol Radiother. 2017 Jan-Feb;22(1):1-9. doi: 10.1016/j.rpor.2016.09.002. Epub 2016 Oct 17.
3
Comparative Evaluation of a 6MV Flattened Beam and a Flattening Filter Free Beam for Carcinoma of Cervix – IMRT Planning Study.
Asian Pac J Cancer Prev. 2018 Mar 27;19(3):639-643. doi: 10.22034/APJCP.2018.19.3.639.
4
5
Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma.
Rep Pract Oncol Radiother. 2017 Jan-Feb;22(1):10-18. doi: 10.1016/j.rpor.2016.09.010. Epub 2016 Oct 17.
6
Volumetric-modulated arc therapy vs. c-IMRT in esophageal cancer: a treatment planning comparison.
World J Gastroenterol. 2012 Oct 7;18(37):5266-75. doi: 10.3748/wjg.v18.i37.5266.
7
The effect of photon energy on intensity-modulated radiation therapy (IMRT) plans for prostate cancer.
Radiat Oncol J. 2012 Mar;30(1):27-35. doi: 10.3857/roj.2012.30.1.27. Epub 2012 Mar 31.
8
Effect of photon-beam energy on VMAT and IMRT treatment plan quality and dosimetric accuracy for advanced prostate cancer.
Strahlenther Onkol. 2011 Dec;187(12):792-8. doi: 10.1007/s00066-011-1150-0. Epub 2011 Nov 29.
10

引用本文的文献

1
The incidence of inflammation among patients suffering from cervix cancer with positive beta haemolytic streptococci cultures from genital tract.
Rep Pract Oncol Radiother. 2019 Sep-Oct;24(5):428-431. doi: 10.1016/j.rpor.2019.07.004. Epub 2019 Sep 11.
2
Feasibility of intensity-modulated radiotherapy to treat gastric cancer.
Rep Pract Oncol Radiother. 2019 Jan-Feb;24(1):68-73. doi: 10.1016/j.rpor.2018.10.006. Epub 2018 Nov 16.
4
Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma.
Rep Pract Oncol Radiother. 2017 Jan-Feb;22(1):10-18. doi: 10.1016/j.rpor.2016.09.010. Epub 2016 Oct 17.
5
Dosimetric analysis of Co-60 source based high dose rate (HDR) brachytherapy: A case series of ten patients with carcinoma of the uterine cervix.
Rep Pract Oncol Radiother. 2016 May-Jun;21(3):201-6. doi: 10.1016/j.rpor.2016.01.003. Epub 2016 Mar 2.
7
Integral dose: Comparison between four techniques for prostate radiotherapy.
Rep Pract Oncol Radiother. 2014 Nov 18;20(2):99-103. doi: 10.1016/j.rpor.2014.10.010. eCollection 2015 Mar-Apr.
8
Complication of radiation therapy among patients with positive S. aureus culture from genital tract.
Rep Pract Oncol Radiother. 2012 May 22;17(4):207-10. doi: 10.1016/j.rpor.2012.03.012. eCollection 2012.
9
Evaluation of combining bony anatomy and soft tissue position correction strategies for IMRT prostate cancer patients.
Rep Pract Oncol Radiother. 2012 Feb 9;17(2):104-9. doi: 10.1016/j.rpor.2012.01.005. eCollection 2012.
10
The importance of accurate treatment planning, delivery, and dose verification.
Rep Pract Oncol Radiother. 2012 Mar 6;17(2):63-5. doi: 10.1016/j.rpor.2012.02.001. eCollection 2012.

本文引用的文献

1
A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy.
Radiother Oncol. 2008 Nov;89(2):180-91. doi: 10.1016/j.radonc.2008.06.013. Epub 2008 Aug 8.
2
Estimation of setup uncertainty using planar and MVCT imaging for gynecologic malignancies.
Int J Radiat Oncol Biol Phys. 2008 Aug 1;71(5):1511-7. doi: 10.1016/j.ijrobp.2008.03.070. Epub 2008 Jun 4.
3
7
Treatments of exceptionally large prostate cancer patients with low-energy intensity-modulated photons.
J Appl Clin Med Phys. 2006 Nov 28;7(4):43-9. doi: 10.1120/jacmp.v7i4.2263.
8
High-energy photons in IMRT: uncertainties and risks for questionable gain.
Technol Cancer Res Treat. 2007 Apr;6(2):147-9. doi: 10.1177/153303460700600212.
9
Interfractional variations in patient setup and anatomic change assessed by daily computed tomography.
Int J Radiat Oncol Biol Phys. 2007 Jun 1;68(2):581-91. doi: 10.1016/j.ijrobp.2006.12.024. Epub 2007 Feb 27.
10
Intensity-modulated radiation therapy, protons, and the risk of second cancers.
Int J Radiat Oncol Biol Phys. 2006 May 1;65(1):1-7. doi: 10.1016/j.ijrobp.2006.01.027.