Scripps Translational Science Institute, La Jolla, Calif.
Fought Statistical Consulting, Chicago, Ill.
Am J Med. 2014 Jan;127(1):95.e11-7. doi: 10.1016/j.amjmed.2013.10.003. Epub 2013 Oct 15.
Cardiac arrhythmias are remarkably common and routinely go undiagnosed because they are often transient and asymptomatic. Effective diagnosis and treatment can substantially reduce the morbidity and mortality associated with cardiac arrhythmias. The Zio Patch (iRhythm Technologies, Inc, San Francisco, Calif) is a novel, single-lead electrocardiographic (ECG), lightweight, Food and Drug Administration-cleared, continuously recording ambulatory adhesive patch monitor suitable for detecting cardiac arrhythmias in patients referred for ambulatory ECG monitoring.
A total of 146 patients referred for evaluation of cardiac arrhythmia underwent simultaneous ambulatory ECG recording with a conventional 24-hour Holter monitor and a 14-day adhesive patch monitor. The primary outcome of the study was to compare the detection arrhythmia events over total wear time for both devices. Arrhythmia events were defined as detection of any 1 of 6 arrhythmias, including supraventricular tachycardia, atrial fibrillation/flutter, pause greater than 3 seconds, atrioventricular block, ventricular tachycardia, or polymorphic ventricular tachycardia/ventricular fibrillation. McNemar's tests were used to compare the matched pairs of data from the Holter and the adhesive patch monitor.
Over the total wear time of both devices, the adhesive patch monitor detected 96 arrhythmia events compared with 61 arrhythmia events by the Holter monitor (P < .001).
Over the total wear time of both devices, the adhesive patch monitor detected more events than the Holter monitor. Prolonged duration monitoring for detection of arrhythmia events using single-lead, less-obtrusive, adhesive-patch monitoring platforms could replace conventional Holter monitoring in patients referred for ambulatory ECG monitoring.
心律失常非常常见,且通常未被诊断,因为它们往往是短暂的且无症状的。有效的诊断和治疗可以显著降低与心律失常相关的发病率和死亡率。Zio 贴片(iRhythm Technologies,Inc,旧金山,加利福尼亚州)是一种新型的、单导联心电图(ECG)、轻便的、获得美国食品和药物管理局批准的、连续记录的可移动粘性贴片监测器,适用于检测接受可移动 ECG 监测的患者的心律失常。
共有 146 名因心律失常评估而就诊的患者同时接受了常规 24 小时动态 Holter 监测和 14 天可移动贴片监测。该研究的主要结果是比较两种设备在总佩戴时间内检测心律失常事件的情况。心律失常事件定义为检测到任何 1 种 6 种心律失常之一,包括室上性心动过速、心房颤动/扑动、大于 3 秒的停顿、房室传导阻滞、室性心动过速或多形性室性心动过速/心室颤动。采用 McNemar 检验比较 Holter 和粘性贴片监测器的配对数据。
在两种设备的总佩戴时间内,粘性贴片监测器检测到 96 次心律失常事件,而 Holter 监测器检测到 61 次心律失常事件(P <.001)。
在两种设备的总佩戴时间内,粘性贴片监测器检测到的事件多于 Holter 监测器。使用单导联、侵入性较小的粘性贴片监测平台进行延长时间监测以检测心律失常事件,可替代接受可移动 ECG 监测的患者的常规 Holter 监测。