The quality of ECG data acquisition, and diagnostic performance of a novel adhesive patch for ambulatory cardiac rhythm monitoring in arrhythmia detection.

作者信息

Karaoğuz M Remzi, Yurtseven Ece, Aslan Gamze, Deliormanlı Bilgen Gülşen, Adıgüzel Ömer, Gönen Mehmet, Li Ko-Mai, Yılmaz Elif Nur

机构信息

Department of Cardiology, Koç University Hospital, Istanbul, Turkey.

Department of Cardiology, Koç University Hospital, Istanbul, Turkey.

出版信息

J Electrocardiol. 2019 May-Jun;54:28-35. doi: 10.1016/j.jelectrocard.2019.02.012. Epub 2019 Feb 28.

Abstract

BACKGROUND

Short and long ambulatory electrocardiographic monitoring with different systems is a widely used method to detect cardiac arrhythmias. In this study, we aimed to evaluate the effectiveness of a novel monitoring device on cardiac arrhythmia detection.

METHODS

We used two different protocols to evaluate device performance. For the first one, 36 healthy subjects were enrolled. The standard 12‑lead, 24-h Holter monitoring and the novel single lead electrocardiogram (ECG) Patch Monitor (EPM) device (BeyondCare®, Rooti Labs Ltd., Taipei, Taiwan) were simultaneously applied to all subjects for 24 h. The quality of ECG data acquisition of novel system was compared to that of standard Holter. The second phase included 73 patients that were referred from our outpatient arrhythmia clinic for evaluation of their symptoms relevant to the cardiac arrhythmias. Advanced algorithms, statistical methods (cross-correlation method, Pearson's correlation coefficient, Bland-Altman plots) were used to process and verify the acquired data.

RESULTS

The overall average beat per minute correlation between BeyondCare® and standard 12‑lead Holter was found 98% in 33 healthy subjects. The mean percentage of invalid measurements in BeyondCare® was 1.6% while the Holter's was 1.7%. In the second protocol of the study, prospective data from 67 patients who were referred for evaluation of their symptoms relevant to cardiac arrhythmias, showed that the mean BeyondCare® wear time was 4.7 ± 0.5 days out of five total days per protocol. The mean analyzable wear time was 93.6%. The water-resistant design enabled 73.5% of the participants to take a shower. 7.3% of participants had minor skin irritations related to the electrodes. Among the patients with detected arrhythmia (40.2% of all patients), 29.6% had their first arrhythmia after the initial two days period. A clinically significant pause was detected in one patient, ventricular tachycardia was detected in four patients, and supraventricular tachycardia was detected in 15 patients. Paroxysmal atrial fibrillation was identified in seven patients. Three of them had their first episodes after the second day of monitoring.

CONCLUSION

BeyondCare® Patch was well-tolerated and allowed prolonged time periods for continuous ECG monitoring, may result in an improvement in clinical accuracy and detection of arrhythmias by cloud-based artificial intelligence operating system.

摘要

背景

使用不同系统进行短期和长期动态心电图监测是检测心律失常的一种广泛应用的方法。在本研究中,我们旨在评估一种新型监测设备在心律失常检测方面的有效性。

方法

我们使用两种不同的方案来评估设备性能。第一种方案中,招募了36名健康受试者。将标准的12导联、24小时动态心电图监测与新型单导联心电图(ECG)贴片监测仪(BeyondCare®,Rooti Labs Ltd.,台北,台湾)同时应用于所有受试者24小时。将新型系统的心电图数据采集质量与标准动态心电图进行比较。第二阶段包括73名从我们的门诊心律失常诊所转诊来评估与心律失常相关症状的患者。使用先进算法、统计方法(互相关法、皮尔逊相关系数、布兰德 - 奥特曼图)来处理和验证采集到的数据。

结果

在33名健康受试者中,发现BeyondCare®与标准12导联动态心电图之间每分钟心跳的总体平均相关性为98%。BeyondCare®中无效测量的平均百分比为1.6%,而动态心电图为1.7%。在研究的第二个方案中,67名转诊来评估与心律失常相关症状的患者的前瞻性数据显示,每个方案的五天中,BeyondCare®的平均佩戴时间为4.7±0.5天。平均可分析佩戴时间为93.6%。防水设计使73.5%的参与者能够洗澡。7.3%的参与者有与电极相关的轻微皮肤刺激。在检测到心律失常的患者中(占所有患者的40.2%),29.6%在最初两天后首次出现心律失常。在一名患者中检测到临床上有意义的停搏,在四名患者中检测到室性心动过速,在15名患者中检测到室上性心动过速。在七名患者中识别出阵发性心房颤动。其中三名在监测第二天后首次发作。

结论

BeyondCare®贴片耐受性良好,允许长时间连续心电图监测,基于云的人工智能操作系统可能会提高临床准确性和心律失常检测率。