Vandenberghe Bart, Jacobs Reinhilde
Oral Imaging Center, Department of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven, Leuven Belgium.
J Oral Maxillofac Res. 2010 Apr 1;1(1):e5. doi: 10.5037/jomr.2010.1105. eCollection 2010.
The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor.
A digital photostimulable storage phosphor (PSP) sensor (Vistascan) and a multipulse X-ray generator (Prostyle Intra) with two tube voltages were used in this study. The front, premolar and molar region of two adult human cadaver skulls jaws were imaged using the X-ray tube at 63 kV and 70 kV, both at 8 mA and decreasing exposure times (160 ms, 120 ms and 80 ms). A standardized exposure protocol containing waxed occlusal keys and an aiming device ensured proper and reproducible beam alignment. Three observers assessed the digital radiographs for 31 selected periodontal bone loss sites. Radiographic measurements were compared to physical measurements (Standard). Subjective ratings of lamina dura, crater defect and furcation involvement visibility, contrast perception and bone quality were also performed.
Multiple regression equation of the variables kV and exposure time demonstrated no significant difference for the periodontal bone level measurements (P > 0.05). In 90.3% and 96.7% of the measurements for 70 kV and 63 kV respectively, deviation was within 1 mm. The subjective ratings produced similar findings in terms of image quality for both tube voltages and the three exposure times.
The results of the present study revealed that tube voltages of 63 kV and 70 kV provided similar accuracy and image quality for periodontal disease diagnosis.
本研究的目的是确定使用数字光激励存储磷光体传感器和两种X射线管电压时,用于牙周疾病诊断的测量准确性和主观图像质量。
本研究使用了数字光激励存储磷光体(PSP)传感器(Vistascan)和具有两种管电压的多脉冲X射线发生器(Prostyle Intra)。使用X射线管在63 kV和70 kV下,均为8 mA并逐渐减少曝光时间(160 ms、120 ms和80 ms),对两个成人尸体颅骨颌骨的前部、前磨牙和磨牙区域进行成像。包含蜡制咬合键和瞄准装置的标准化曝光方案确保了光束对准的正确和可重复性。三名观察者对31个选定的牙周骨丧失部位的数字X线片进行评估。将X线片测量结果与物理测量结果(标准)进行比较。还对硬骨板、牙槽窝缺损和根分叉病变的可见性、对比度感知和骨质进行了主观评分。
变量kV和曝光时间的多元回归方程显示,牙周骨水平测量无显著差异(P>0.05)。在70 kV和63 kV测量中,分别有90.3%和96.7%的偏差在1 mm以内。对于两种管电压和三种曝光时间,主观评分在图像质量方面得出了相似的结果。
本研究结果表明,63 kV和70 kV的管电压在牙周疾病诊断中提供了相似的准确性和图像质量。