Evaluation of a new 3-dimensional color Doppler flow method to quantify flow across the mitral valve and in the left ventricular outflow tract: an in vitro study.

机构信息

Department of Pediatric Cardiology, Oregon Health and Science University, 3181 SW Sam Jackson Park Dr, L608, Portland, OR 97239-3098 USA.

出版信息

J Ultrasound Med. 2014 Feb;33(2):265-71. doi: 10.7863/ultra.33.2.265.

Abstract

OBJECTIVES

The aim of this study was to assess the accuracy, feasibility, and reproducibility of determining stroke volume from a novel 3-dimensional (3D) color Doppler flow quantification method for mitral valve (MV) inflow and left ventricular outflow tract (LVOT) outflow at different stroke volumes when compared with the actual flow rate in a pumped porcine cardiac model.

METHODS

Thirteen freshly harvested pig hearts were studied in a water tank. We inserted a latex balloon into each left ventricle from the MV annulus to the LVOT, which were passively pumped at different stroke volumes (30-80 mL) using a calibrated piston pump at increments of 10 mL. Four-dimensional flow volumes were obtained without electrocardiographic gating. The digital imaging data were analyzed offline using prototype software. Two hemispheric flow-sampling planes for color Doppler velocity measurements were placed at the MV annulus and LVOT. The software computed the flow volumes at the MV annulus and LVOT within the user-defined volume and cardiac cycle.

RESULTS

This novel 3D Doppler flow quantification method detected incremental increases in MV inflow and LVOT outflow in close agreement with pumped stroke volumes (MV inflow, r = 0.96; LVOT outflow, r = 0.96; P < .01). Bland-Altman analysis demonstrated overestimation of both (MV inflow, 5.42 mL; LVOT outflow, 4.46 mL) with 95% of points within 95% limits of agreement. Interobserver variability values showed good agreement for all stroke volumes at both the MV annulus and LVOT.

CONCLUSIONS

This study has shown that the 3D color Doppler flow quantification method we used is able to compute stroke volumes accurately at the MV annulus and LVOT in the same cardiac cycle without electrocardiographic gating. This method may be valuable for assessment of cardiac output in clinical studies.

摘要

目的

本研究旨在评估一种新型的 3 维(3D)彩色多普勒流量定量方法测量二尖瓣(MV)流入和左心室流出道(LVOT)流出时不同心排量的准确性、可行性和可重复性,与在水泵猪心模型中的实际流量相比。

方法

我们将乳胶气球从 MV 瓣环插入到 LVOT 中,在水槽中对 13 个新鲜收获的猪心进行了研究,通过校准的活塞泵以 10ml 的增量将其在不同的心排量(30-80ml)下被动泵入,未进行心电图门控的 4 维流量测量。使用原型软件对数字成像数据进行离线分析。在 MV 瓣环和 LVOT 放置了两个半球形的彩色多普勒速度测量采样平面。软件在用户定义的体积和心动周期内计算 MV 瓣环和 LVOT 的流量。

结果

这种新型的 3D 多普勒流量定量方法检测到 MV 流入和 LVOT 流出呈递增趋势,与泵送的心排量密切一致(MV 流入,r=0.96;LVOT 流出,r=0.96;P<.01)。Bland-Altman 分析表明,两种方法均存在高估(MV 流入,5.42ml;LVOT 流出,4.46ml),95%的点在 95%一致性界限内。在 MV 瓣环和 LVOT 处,所有心排量的观察者间变异性值均显示出良好的一致性。

结论

本研究表明,我们使用的 3D 彩色多普勒流量定量方法能够在同一心动周期内准确地计算 MV 瓣环和 LVOT 的心排量,无需心电图门控。该方法在临床研究中评估心输出量可能具有重要价值。