Bezuti Márcio Takey, Mandarano Luiz Garcia, Barbieri Giuliano, Mazzer Nilton, Barbieri Cláudio Henrique
Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo - Ribeirão Preto, SP, Brazil.
Acta Ortop Bras. 2013 Jan;21(1):46-51. doi: 10.1590/S1413-78522013000100010.
To measure the ultrasound propagation velocity (UV) through a tibial transverse osteotomy in sheep, before and after the fixation with a DCP plate.
Ten assemblies of a DCP plate with the diaphyseal segment of tibiae, in which a transverse osteotomy was made, were used. Both coronal and sagittal transverse and the axial UV were measured, first with the intact bone assembled with the plate and then with the uncompressed and compressed osteotomy; statistical comparisons were made at the 1% (p<0.01) level of significance.
Compared with the intact bone assembly, axial UV significantly decreased with the addition of the osteotomy and significantly increased with compression, presenting the same behavior for the other modalities, although not significantly.
In accordance with the literature data on the ultrasonometric evaluation of fracture healing, underwater UV measurement was able to demonstrate the efficiency of DCP plate fixation. The authors conclude that the method has a potential for clinical application in the postoperative follow-up of DCP plate osteosinthesis, with a capability to demonstrate when it becomes ineffective. Laboratory investigation.