In vivo standardization of bone ultrasonometry of the clavicle.

作者信息

Mandarano-Filho Luiz Garcia, Bezuti Márcio Takey, Barbieri Cláudio Henrique

机构信息

Hospital das Clínicas, Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ortopedia, Ribeirão Preto/SP, Brazil.

出版信息

Clinics (Sao Paulo). 2016 Mar;71(3):140-4. doi: 10.6061/clinics/2016(03)04.

Abstract

OBJECTIVE

The assessment of fracture union includes physical examination and radiographic imaging, which depend on the examiner's experience. The development of ancillary methods may avoid prolonged treatments and the improper removal of implants. Quantitative bone ultrasonometry has been studied for this purpose and will soon be included in clinical practice. The aims of the present study were to assess the feasibility of using this technique on the clavicle and to standardize its in vivo application.

METHODS

Twenty adult volunteers, including 10 men and 10 women without medical conditions or a previous history of clavicle fracture, underwent axial quantitative ultrasonometric assessment using transducers in various positions (different distances between the transducers and different angulations relative to the clavicle).

RESULTS

Similar values of wave propagation velocity were obtained in the different tested set-ups, which included distinct distances between the transducers and angular positions relative to the clavicle. There were significant differences only in the transducers positioned at 0° and at 5 or 7 cm apart.

CONCLUSIONS

The use of bone ultrasonometry on the clavicle is feasible and the standardization of the technique proposed in this study (transducers placed at 45° and at 7 cm apart) will allow its future application in clinical trials to evaluate the healing process of diaphyseal fractures of the clavicle.

摘要

目的

骨折愈合的评估包括体格检查和影像学检查,这取决于检查者的经验。辅助方法的发展可能避免延长治疗时间和不当取出植入物。为此已经对定量骨超声检查进行了研究,并且该检查很快将被纳入临床实践。本研究的目的是评估在锁骨上使用该技术的可行性并规范其在体内的应用。

方法

20名成年志愿者,包括10名男性和10名女性,无疾病史或既往锁骨骨折史,使用换能器在不同位置(换能器之间的不同距离以及相对于锁骨的不同角度)进行轴向定量超声评估。

结果

在不同的测试设置中获得了相似的波传播速度值,这些设置包括换能器之间不同的距离以及相对于锁骨的不同角度位置。仅在相距5或7厘米且角度为0°的换能器位置存在显著差异。

结论

在锁骨上使用骨超声检查是可行的,并且本研究中提出的技术标准化(换能器以45°角放置且相距7厘米)将使其未来能够应用于临床试验,以评估锁骨骨干骨折的愈合过程。