Eur J Heart Fail. 2014 Apr;16(4):403-8. doi: 10.1002/ejhf.52.
Continuous-flow left ventricular assist device (CF-LVAD) implantation is associated with improved quality of life, but the effect on exercise capacity is less well documented. It is uncertain whether a fixed CF-LVAD pump speed, which allows for sufficient circulatory support at rest, remains adequate during exercise. The aim of this study was to evaluate the effects of fixed versus incremental pump speed on peak oxygen uptake (peak VO2) during a maximal exercise test.
In CF-LVAD (HeartMate II) patients exercise testing measuring peak oxygen uptake (VO2) was performed on an ergometer bike twice in one day: once with fixed pump speed (testfix) and once with incremental pump speed (testinc). The order of testfix and testinc in each patient was determined by randomization. During testinc pump speed was increased from the baseline value by 400 rpm/2 min. Fourteen patients (aged 23–69 years) were included with a mean support duration of 465±483 days. Baseline CF-LVAD speed was 9357±238 rpm and during testinc speed was increased by a mean of 1486±775 rpm. Mean peak VO2 was significantly higher in testinc compared with testfix (15.4±5.9 mL/kg/min vs. 14.1±6.3 mL/kg/min; P=0.012), corresponding to a 9.2% increase. All exercise tests (n=28) were adequately performed with RER>1.
Increasing pump speed during exercise augments peak VO2 in patients supported with CF-LVADs. An automatic speed-change function in future generations of CF-LVADs might improve functional capacity.
持续流左心室辅助装置(CF-LVAD)的植入与生活质量的改善相关,但运动能力的改善效果记录较少。在休息时能提供足够循环支持的固定 CF-LVAD 泵速在运动时是否仍然足够,这一点尚不确定。本研究的目的是评估固定与递增泵速对最大运动测试中峰值摄氧量(peak VO2)的影响。
在 CF-LVAD(HeartMate II)患者中,在一天内两次在跑步机上进行运动测试以测量峰值摄氧量(VO2):一次采用固定泵速(testfix),一次采用递增泵速(testinc)。每位患者的 testfix 和 testinc 的顺序由随机确定。在 testinc 中,泵速从基线值以 400rpm/2min 的速度增加。共纳入 14 名(年龄 23-69 岁)患者,平均支持时间为 465±483 天。基线 CF-LVAD 速度为 9357±238rpm,在 testinc 中速度增加了 1486±775rpm。与 testfix 相比,testinc 中的平均峰值 VO2 明显更高(15.4±5.9mL/kg/min 比 14.1±6.3mL/kg/min;P=0.012),对应增加了 9.2%。所有 28 次运动测试(n=28)均以 RER>1 充分进行。
在 CF-LVAD 支持的患者中,在运动时增加泵速可增加峰值 VO2。未来几代 CF-LVAD 中的自动速度变化功能可能会提高功能能力。