Cobelo-García Antonio, Santos-Echeandía Juan, López-Sánchez Daniel E, Almécija Clara, Omanović Dario
Grupo de Bioxeoquímica Mariña, Instituto de Investigacións Mariñas (IIM-CSIC) , Vigo, Pontevedra, Spain.
Anal Chem. 2014 Mar 4;86(5):2308-13. doi: 10.1021/ac403558y. Epub 2014 Feb 13.
The determination of trace elements using stripping voltammetry may be seriously affected by the presence of intensive matrix background or interfering peaks, leading to poorer detection limits and/or inaccurate quantitative results. In this work, we have tested the use of signal transformation (e.g., second derivative) in the analysis of platinum in seawater and sediment digests by means of catalytic adsorptive stripping voltammetry. In natural waters, the limit of detection of Pt is affected by a broad background wave due to the formazone complex used in the sample matrix for its determination, while in sediment digests, the Pt peak may be interfered with due to the presence of elevated concentrations of Zn, affecting the accuracy of the determination. Results applying second derivative signal transformation revealed a significant improvement (2-3-fold) of the detection limit in water due to the minimization of background effects, therefore allowing shorter accumulation times and faster determinations. In the presence of interfering peaks, the inaccuracy resulting from erroneous baseline selection in the original signal is eliminated when the second derivative is used. Signal processing should be considered as a useful tool for other voltammetric methodologies where more accurate or faster determinations are needed.
使用溶出伏安法测定微量元素时,强烈的基体背景或干扰峰的存在可能会严重影响测定结果,导致检测限变差和/或定量结果不准确。在本研究中,我们通过催化吸附溶出伏安法测试了信号转换(如二阶导数)在海水和沉积物消解液中铂分析中的应用。在天然水体中,由于用于样品基体中铂测定的甲臜配合物会产生宽背景波,铂的检测限受到影响;而在沉积物消解液中,由于锌浓度升高,铂峰可能会受到干扰,影响测定的准确性。应用二阶导数信号转换的结果表明,由于背景效应最小化,水中的检测限有显著提高(2至3倍),因此可以缩短富集时间并加快测定速度。在存在干扰峰的情况下,使用二阶导数时可消除原始信号中错误基线选择导致的不准确。对于需要更准确或更快测定的其他伏安法,信号处理应被视为一种有用的工具。